Selecting the measurement data type
Select the measurement data type by pressing the ACC/VEL/DISP key.
Acceleration, Velocity, Displacement, and Envelope are selectable.
③
Input range setting
Set the input range to an optimal setting. The optimal setting is to use the lowest range where an overload does
not occur. The key changes to the higher range, the key changes to the lower range.
By changing input range, an electric transient noise occurs and overload may be displayed.
In t
his case, it is
not an excessive input of vibration. When an overload is displayed immediately after changing input range,
please wait for a while. The overload may disappear.
When the overload occurs, the “Over” indication appears to the left side of display.
ACC/VEL/DISP/
ENVELOPE
Frequency span
Number of analysis
lines
Upper limit time
ACC/VEL/DISP key
RANGE key
Overload state
Overload indication