Rion VA-12 Quick Instruction Manual Download Page 12

 

12 

 

number of analysis lines. If the generated frequency cannot be predicted, measure the frequency roughly with 

high  frequency  span  at  first,  then  set  the  frequency  span  and  the  number  of  analysis  lines  so  that  the 

frequency resolution becomes narrower than 1 Hz. 

 

  Setting the Operation type 

In  the  Instantaneous  measurement,  it  is  not  necessary  to  set  the  operation  type.  In  the  measurement  with 

operation  function,  it  is  necessary  to  select  the  operation  type.  Then  press the  INST/AVE  key,  and  press the 

START/STOP key. 

In order to select the operation type, presses the MENU key, then go into the submenu [FFT] from the Menu 

List, and set the Operation type and the Average number. 

 

 

 

 

 

 

 

 

 

 

  Instantaneous measurement 

FFT spectra are continuously  displayed without averaging. It is used  when grasping the  characteristics of 

phenomenon by observing the time variation  of the spectrum,  or when the average is not  necessary since 

the vibration is steady. 

 

The operation type is chosen as follows according to the measurement purpose. 

◆ 

Linear average 

It  is  a  popular  averaging  type,  and  is  used  to  obtain  the  arithmetic  averaged  result.  The  operation  stops 

when the number of measured data reaches to the Average number set in the submenu [FFT]. 

◆ 

Exponential average 

FFT  spectra  are  continuously  displayed  with  a  time  constant.  When  amplitude  of  vibration  changes 

violently,  the  averaged  amplitude  becomes  smooth.  It  is  used  when  grasping  the  characteristics  of 

phenomenon by  observing  the time variation  of the spectrum.  If an impulsive vibration occurs,  you  may 

miss it in Instantaneous measurement. However you will not miss it in exponential average, since the time 

constant keeps the influence of the impulsive vibration for a while. 

◆ 

Maximum hold 

It is used when obtaining the maximum value of each frequency component. The operation stops when the 

number of measured data reaches to the Average number set in the submenu [FFT]. 

Setting the Operation type 

Setting the Average number 

Summary of Contents for VA-12

Page 1: ...ion 2 Measurement in Vibration Meter Mode 3 Measurement in Time Waveform Mode 4 Measurement in FFT Analyzer Mode 5 Examples of typical vibration measurement 5 1 Diagnosis of unbalance and misalignment...

Page 2: ...sis functions Vibration Meter Mode Time Waveform Mode and FFT Analyzer Mode The operation about common setups to each analysis function before starting measurement is explained in this section Inserti...

Page 3: ...7I Other accelerometers than PV 57I Optional accessory When the amplitude of vibration is too small or too large or the frequency is too high the PV 57I cannot measure them In that case an acceleromet...

Page 4: ...ching the accelerometer it is necessary to wait about 30 seconds until the influence of the impulsive vibration disappears Moreover it is necessary to care about the contact resonance frequency of the...

Page 5: ...selectable Input range setting Set the input range to the optimal setting The optimal setting is to use the lowest range where an overload does not occur The key changes to the higher range the key ch...

Page 6: ...ured data Store the data if necessary When the STORE key is pressed the data at that time is stored on SD card in VA 12 When in a paused state the fixed data is stored 3 Measurement in Time Waveform M...

Page 7: ...igher range the key changes to the lower range By changing input range an electric transient noise occurs and overload may be displayed In this case it is not an excessive input of vibration When an o...

Page 8: ...key The measurement changes to the paused state and a measured time waveform is fixed To resume the measurement press the PAUSE CONT key one more time Zooming the X axis and the Y axis of the graph d...

Page 9: ...STORE key is pressed the data at that time is stored on SD card in VA 12 When in a paused state the fixed data is stored 4 Measurement in FFT Mode Selecting the analysis function to FFT mode after com...

Page 10: ...hout repetition like a single pulse Hanning It is used for the usual measurement Flattop It is used for reading the amplitude correctly although the frequency resolution may be worse Selecting the mea...

Page 11: ...y span and the Number of analysis lines In order to determine the maximum frequency to analyze and the frequency resolution the frequency span and the number of analysis lines are set up The frequency...

Page 12: ...e average is not necessary since the vibration is steady The operation type is chosen as follows according to the measurement purpose Linear average It is a popular averaging type and is used to obtai...

Page 13: ...rum In Linear average or Maximum hold the operation stops when the average counter reaches to the average number and the FFT spectrum is fixed In Instantaneous measurement or Exponential average the o...

Page 14: ...is assumed that the cursor move symbol is displayed at first If the Graph zoom symbol is displayed at first press the SCALE EXPAND MOVE key to change the zoom mode Move the cursor to the position whe...

Page 15: ...d in VA 12 When in a paused state the fixed data is stored STORE key PAUSE CONT key Original graph Move the cursor to the position you want to expand Expand 4 Expand 8 Expand 16 Cursor Cursor SCALE EX...

Page 16: ...are investigated by Time Waveform Mode or FFT Analyzer Mode In unbalance condition only a shaft rotational frequency appears but in misalignment condition a shaft rotational frequency and its harmonic...

Page 17: ...udgment is abnormal the precise diagnosis is carried out In precise diagnosis the defected part can be predicted by FFT analysis of acceleration envelope if the information about the bearing to diagno...

Page 18: ...of VA 12 about the details of the trigger function When an exciting point and an attached point of the accelerometer changes to the other position the measured frequencies may change In this case all...

Page 19: ...e trigger ON Trigger level 1 8 2 8 Slope The trigger conditions are set in the submenu Measurement Press the MENU key and go into the submenu Measurement from the Menu List A measurement example of na...

Page 20: ...ddress is indicated at the upper right corner of display This store address means the address contained in a Store Name which was set up at the previous step The measured data can be stored in the ind...

Page 21: ...e the selected data is recalled and the character Recall is displayed at the upside center The recall address can be selected by keys Returning to the measurement state The START STOP key is pressed t...

Page 22: ...reen bitmap image Press both the MENU key and the ENT key simultaneously When the storing is started the indicator blinks red It disappears after completed Recalling the screen bitmap image Press both...

Page 23: ...s the SD card in VA 12 as a removable disk All stored files exist in the VA 12 folder The folders MAN_ include measured data files means 4 digits number The folder name MAN_ is same as the Store Name...

Page 24: ...following figure there are six data files in the folder MAN_9999 The extension names of the data files are RVD Copy required data files to another folder in a computer Copying BMP files The stored BM...

Page 25: ...The file open window appears as shown in the following figure In this window select the folder to import for example MAN_9999 In this stage since the file type is set to All Excel Files No items match...

Page 26: ...ining that the selected file is delimited After clicking the Open the Text Import Wizard Step 1 of 3 appears as shown in the following figure In this wizard select the Choose the file type that best d...

Page 27: ...ile After clicking the Next the Text Import Wizard Step 2 of 3 appears Check the checkbox of the Comma Then click the Finish Importing the data to Microsoft Excel After clicking the Finish the data ar...

Reviews: