5. How to Operate
5-3. How to activate the gas detector
- 27 -
<Start-up Procedures (approximately 25 seconds for system check of the gas detector and alarm
deactivation)>
Power On
↓
PW
A1
A2
F
LCD
Initial clear
●
○
○
○
----
WARM UP
↓
↓
Gas Specifications Display
●
○
○
○
15.0 ppm
COS
↓
↓
Detection mode
●
○
○
○
0.0 ppm
COS
CAUTION
Do not turn off the gas detector during the initial clear. The gas detector is reading the sensor
memory during the initial clear.
When the power is turned off to perform replacement or maintenance of parts, the gas detector
needs to be warmed up for a specified period which is determined depending on the length of
unpowered time.
During the warm-up, the alarm activation and output signals are unstable. Provide a prior
notification to the related sections so that they can prepare for false abnormalities.
After the warm-up is completed, check that the reading on the flow rate indicator corresponds to
the specified flow rate, and then perform a calibration.