Riken Keiki RP-D58 AC Operating Manual Download Page 50

10 Product Specifications

 

10-2. Detection principle 

 

RP-D58 Series

 

- 48 - 

 

10-2. Detection principle 

 

[Flow Sensor] 
The flow sensor has a circuit that can output a contact signal when the flow rate drops below the preset 
level. 
The flow sensor is a hot-wire type with a Wheatstone bridge consisting of a detecting element, 
compensating element, and fixed resistance. Contained in a flame arrester container allows it to be 
flame-proof.  
When the drawn air or gas comes into contact with the heated detecting element, the detecting element 
temperature changes due to the heat dissipation, and the electrical resistance of the element changes 
according to the temperature changes. Since the resistance changes are in proportion to the gas flow rates, 
the Wheatstone bridge circuit can identify the changes as the flow rate. 
 

 

 

 

 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Structure Diagram 

D: Detecting element 

C: Compensating   

element 

Fixed 

resistance 

Fixed 

resistance 

Variable resistor for bridge 

equilibrium adjustment 

Variable resistor for bridge 

voltage adjustment 

Switch 

Power supply 

Basic Circuit 

VR

SW 

Flow Sensor Appearance 

Summary of Contents for RP-D58 AC

Page 1: ...PT9E 0128 Fixed type Explosion Proof Pump Unit RP D58 AC RP D58 DC Operating Manual...

Page 2: ...unit 16 4 2 Precautions for installation points 16 4 3 Precautions for system designing 17 4 4 How to install 19 4 5 How to wire 21 4 6 How to tube 29 5 How to Operate 30 5 1 Preparation for start up...

Page 3: ...led into the atmosphere and sending into a gas detector head This pump unit is used combining the indicator alarm unit and the gas detector head if needed The pump unit can be used by itself in order...

Page 4: ...TE This message indicates advice on handling 1 4 Method of confirmation for Standards and Explosion proof specification This instrument has some specification depends on standard and explosion proof c...

Page 5: ...pes liquid or vapor It may cause the color and shape of the window plate to be changed The flameproof joints are not intended to be repaired Do not open the lid when applying current The lid may be op...

Page 6: ...otective functions for defects When seeming defects are found in the protective functions such as protective grounding do not start the pump unit External connection Before connecting the pump unit to...

Page 7: ...the settings without understanding the specifications may cause alarm malfunctions Please use the pump unit properly in accordance with the operating manual Avoid applying organic solvents and others...

Page 8: ...uble ATEX Specifications Technical Data Protection Method Flameproof enclosure d Group II Category 2G Type of Protection and Marking code Ex db B H2 T4 Equipment Protection Level Gb Ambient Temperatur...

Page 9: ...53 C Electrical Data Supply voltage Type AC AC100 110V 50 60Hz 120mA Type DC DC24V 360mA Contact output AC250V 0 5A Load resistance DC30V 0 5A Load resistance Applicable Standard JNIOSH TR NO 43 2008...

Page 10: ...C Electrical Data Supply voltage Type AC AC100 110V 50 60Hz 120mA Type DC DC24V 360mA Contact output AC250V 0 5A Load resistance DC30V 0 5A Load resistance Applicable Standard JNIOSH TR NO 43 2008 In...

Page 11: ...standard accessories 9 RP D58 Series 3 Product Components 3 1 Main unit and standard accessories Main Unit including cable glands Unit mm Gas outlet Display Gas inlet Cable connecting port for pump p...

Page 12: ...number of wrenches as with the test certificates will be provided Dust removal filter with flow monitor 1 filter NOTE The dedicated control key is for maintenance It is not used for detection mode Ke...

Page 13: ...r change a value DOWN 5 Power lamp Power lamp Detection mode It lights up in green Maintenance mode It blinks in green 6 Fault lamp Fault lamp It lights up in yellow when an abnormality is detected in...

Page 14: ...t lid Remove the electric circuit lid and display unit to find behind them the electric circuit section shown in the figure above Power input terminal plates are RP D58 AC 3 pole and RP D58 DC 2 pole...

Page 15: ...he figure above The bolt with the hexagon socket must use the stainless steel material of property class A2 70 Grease specified by RIKEN KEIKI BARRIERTA JFE 552 manufactured by NOK KLUBER If you can n...

Page 16: ...ries 14 3 3 Block diagram Electric Diagram RP D58 AC Controller CPU POWER OUTPUT 24 VDC AC DC converter Display POWER FAULT Seven segment LED four digit Magnetic operation part MENU ESC SET Flow senso...

Page 17: ...s 3 3 Block diagram 15 RP D58 Series Electric Diagram RP D58 DC Controller CPU Display POWER FAULT Seven segment LED four digit Magnetic operation part MENU ESC SET Flow sensor POWER INPUT 24 VDC Powe...

Page 18: ...a place where the range of operating temperatures is exceeded The pump unit must be installed in a stable place where the operating temperature is maintained and do not change suddenly ATEX Specifica...

Page 19: ...ases exist around it 4 3 Precautions for system designing Using a stable power supply The external output and alarm contact of the pump unit may be activated when the power is turned on when momentary...

Page 20: ...y lightning is impossible the following protective measures can be taken Protection against lightning Take appropriate measures in accordance with the importance of the facilities and the environment...

Page 21: ...mp unit Attach the pump unit on the wall or elsewhere using four M6 screws Tighten the hexagon socket head cap bolt fixing the lid and the main body with a tightening torque of 215 6 24 5N cm When clo...

Page 22: ...ed to perform maintenance or where the pump unit cannot be removed because tubes or racks prevent access to it Do not install the pump unit in a place where maintenance involves dangers for example ne...

Page 23: ...5 17 16 2 No cable Plug WARNING When installing the electric circuit lid do not let any foreign substance such as metal get into the terminal box This can cause a failure of the device or loss of expl...

Page 24: ...ble range 10 5 11 0 11 14 11 8 11 0 allowable range 11 0 11 5 12 14 12 8 11 5 allowable range 11 5 12 0 12 14 12 8 12 0 allowable range 12 0 12 5 12 5 14 12 8 13 0 allowable range 13 0 13 5 13 5 14 13...

Page 25: ...DC 2 pole A fter use closely tighten the lid until the thread is no longer seen clockwise 8 rotations 1 4 rotation or more and the TOP marking faces upward Tighten the hexagon socket set screw with a...

Page 26: ...amping of the wire due to a shorter bare wire length may cause defective electric conduction or heating Catching the wire insulation due to a shorter bare wire length may cause defective electric cond...

Page 27: ...er the wire is connected securely pull the wire gently Do not pull the wire strongly Push the lever with your finger to lower the spring in its inside While holding down the lever insert the wire into...

Page 28: ...e gland and wipe the specified grease CAUTION Tighten a cable gland with a tool until a clearance between the cable gland and a main unit case is below 2 0 mm If it is difficult to tighten the cable g...

Page 29: ...NOK KLUBER If you can not prepare the specified grease use one that meets the following requirements 1 Material does not harden due to deterioration 2 Volatile solvent free 3 Material does not cause c...

Page 30: ...head signal Pump power supply 100 110 VAC 2 1 WARNING Before turning on the pump unit do not forget to connect it to a grounding terminal For stable operation of the pump unit and safety it must be c...

Page 31: ...The pump unit has an Rc1 8 thread inside of the sampling inlet outlet GAS IN GAS OUT to which BS unions are attached as standard Because their material varies depending on the gas to be detected plea...

Page 32: ...unit is connected to a grounding circuit Check that the wiring is connected to external device properly Check that the tubes are installed properly Check that the power supply voltage is compliant wi...

Page 33: ...h Since RP D58 DC does not have any power switch there is no need to open or close the electric circuit lid NOTE Start up Operation Procedures approximately 25 seconds for system check of the pump uni...

Page 34: ...tor alarm unit 2 Only RP D58 AC has a power switch Since RP D58 DC does not have any power switch there is no need to open or close the electric circuit lid WARNING When the pump unit is turned off an...

Page 35: ...immediately Low Flow Rate Abnormal Operations This alarm is triggered when the output value of the low flow sensor drops below the alarm setpoint The alarm is auto reset when the flow rate exceeds a...

Page 36: ...ate detection 7 1 Maintenance intervals and items Daily maintenance Perform maintenance before beginning to work Regular maintenance Perform maintenance once or more every six months to maintain the p...

Page 37: ...flow rate threshold value setting 2 6 3 Flow sensor zero point check 2 6 4 Flow sensor output check 2 6 5 Pump output check 2 6 7 Flow sensor operation start setting 2 6 8 Low flow rate alarm delay ti...

Page 38: ...SET key to perform the zero adjustment 1300 Under zero adjustment CAL is displayed Wait for a while until the adjustment is completed CAL Zero adjustment completed It automatically returns to 2 6 1 a...

Page 39: ...the threshold value for low flow rate to one of the symbols a to e printed on the flow sensor by pressing the or key Flow rate threshold value setting Press the SET key C Once the value is set the dis...

Page 40: ...n start status Select either ON OFF using the UP DOWN key Select ON to start the operation OFF Flow sensor operation start setting Press the SET key ON Once the value is set the display returns to 2 6...

Page 41: ...yed If the zero point check of the flow sensor fails it automatically returns to 2 6 3 after FAIL is displayed PASS Return to 2 6 3 Install the test pump PW A F LED 2 6 4 Flow sensor output check Chec...

Page 42: ...s List of recommended regular replacement parts Name Maintenance intervals Replacement intervals year Quantity pieces unit Remarks AC DC converter 6 months 3 6 years 1 Flow sensor 6 months 2 years 1 T...

Page 43: ...re not present 8 2 Procedures to relocate the pump unit or use it again When the pump unit is relocated select a new place in accordance with 4 2 Precautions for installation points and 4 4 How to ins...

Page 44: ...lities momentary blackout of power supply system Provide the rated voltage Take measures such as checking or adding the UPS power supply line filter and insulation transformer Cable abnormalities open...

Page 45: ...istant load Power contact cable CVVS etc 1 25mm2 or 2 0mm2 2 core or 4 core Transmission distance Less than 375m in case of CVV 1 25mm2 Less than 550m in case of CVV 2 0mm2 Initial clear Approx 25sec...

Page 46: ...or 3 core Transmission distance Less than 1 25km in case of Cable 1 25mm2 Less than 2 0km in case of Cable 2 0mm2 Contact cable CVVS etc 1 25mm2 or 2 0mm2 2 core Initial clear Approx 25sec Power suppl...

Page 47: ...cable CVVS etc 1 25mm2 or 2 0mm2 2 core or 4 core Transmission distance Less than 375m in case of CVV 1 25mm2 Less than 550m in case of CVV 2 0mm2 Initial clear Approx 25sec Power supply 24VDC 10 Powe...

Page 48: ...3 core Transmission distance Less than 1 25km in case of Cable 1 25mm2 Less than 2 0km in case of Cable 2 0mm2 Contact cable CVVS etc 1 25mm2 or 2 0mm2 2 core Initial clear Approx 25sec Power supply...

Page 49: ...ble CVVS etc 1 25mm2 or 2 0mm2 2 core or 4 core Transmission distance Less than 375m in case of CVV 1 25mm2 Less than 550m in case of CVV 2 0mm2 Initial clear Approx 25sec Power supply 24VDC 10 Power...

Page 50: ...rawn air or gas comes into contact with the heated detecting element the detecting element temperature changes due to the heat dissipation and the electrical resistance of the element changes accordin...

Page 51: ...10 Product Specifications 10 2 Detection principle 49 RP D58 Series...

Reviews: