9 Product Specifications
9-1. List of specifications
- 29 -
9
Product Specifications
9-1. List of specifications
Model
GD-1DOX-AS
Detection principle
Galvanic cell method
Detection gas
O
2
Applicable gas detector head
GD-1DOX
Detection method
Aspirator suction type
Sampling condition
Equivalent to atmospheric pressure (compressed air: 0.1 - 0.7 MPa, maximum use
amount: 3 L/min <converted into atmospheric pressure>)
Suction flow
1.5±0.5L/min
Transmission cable
CVVS worth of shield cable(1.25mm
2
or 2.0mm
2
)
・
2-core
Cabling port
*1
Pressure proof packing gland<G3/4>(Compatible cables
φ
9.6
~
φ
13.0mm in outer
diameter)
Piping port
*2
Gas IN: Rc1/4 (O.D ø6-1t, half-union for PTFE tube <PP> supplied)
OUT: O.D ø8-1t, flare union for Cu tube <Bs>
Compressed air IN: O.D ø8-1t, flare union for Cu tube <Bs>
Operating temperature range
-10
~
+40
℃
(non-rapidly-vary)
Operating humidity range
Below 95%RH(non-condensing/non-rapidly-vary)
Structure
Wall mounting sampling unit
Explosion-proof structure
Flame proof structure
Explosion-proof class
Ex d
Ⅱ
C T6 X(TIIS,Japan)
Dimensions
Approx.400(W)×265(H)×102(D)mm(projection portions excluded)
Weight
Approx.6.3kg
Color
Detection body: Munsell 7.5BG5/2
*1 Please specify your request when ordering.
*2 When a half-union for gas IN side is changed, check also the external filter (IN/OUT diameter) carefully.
Summary of Contents for GD-1DOX-AS
Page 1: ...PT2E 2880 Fixed type Oxygen Detector Head GD 1DOX AS Operating Manual PT2 183...
Page 9: ......
Page 10: ......
Page 15: ......
Page 16: ......
Page 18: ......
Page 22: ......
Page 25: ......
Page 32: ......
Page 34: ...Manual Log Rev Amendment Issue data 0 First issue PT2 2880 1 31 2022...