39/60
• Saving settings
Now enter the following in the
④
dialog box:
save
Click
⑤
[Apply] again, then click the
⑥
[Apply] button that appears in the upper menu.
The command is entered and the setting is saved.
• Restarting
Now enter the following in the
④
dialog box:
reboot
Click
⑤
[Apply] again, then click the
⑥
[Apply] button that appears in the upper menu.
The command is entered and the setting is saved.
This completes the procedure for changing the Ewon Flexy 205 (GADMS Lite) IP address.
The gas detector and PC (local connection) IP addresses must also be changed after
changing the Ewon Flexy 205 IP address.
7-3-4
Setting the time (Only when using a local connection)
Double-click the [Value] column for
③
’ [SETTING_STATEMENT].
Enter the following in the
④
dialog box:
time dd/mm/yyyy hh:mm:ss (example: “time 30/01/2020 17:10:30”)
Click
⑤
[Apply], then click the
⑥
[Apply] button that appears in the upper menu. The
command is entered and the time is updated.
When connected to the Internet, the time is set automatically. The time will not be altered
even if this procedure is performed.
7-3-5
Changing the Ewon Flexy 205 (GADMS Lite) login password
• Changing the password for Username: admin (operator rights)
Double-click the [Value] column for
③
’ [SETTING_STATEMENT].
Enter the following in the
④
dialog box:
sysuser admin password ******
Click
⑤
[Apply], then click the
⑥
[Apply] button that appears in the upper menu. The
command is entered and the password is updated.