3 How to Install
3-4. How to Tube
- 19 -
(7)
As a general rule, exhaust should be atmosphere release. When the tip is outside, bend the shape of
the tube into an inverted U to prevent rain from entering the tube. When it returns to the exhaust duct
etc., keep the state within the atmosphere pressure of ±3 kPa, and the state that pressure fluctuation
is not rapidly changed.
(8)
For safety, to provide a flame arrestor for line, install it into each line of the GAS IN and GAS OUT.
CAUTION
FI-915 requires appropriate tubing works (selection of materials and other works) depending on
types of sample gas and installation conditions. For questions about tubing works, please
contact
the distributor.
Summary of Contents for FI-915