18
4-4.
Recommendable spare parts list
No.
Description
Inspection
interval
Replacement
interval
Q’ty/set
1
Sealing packing
(Non-gas contact part)
1 year
7~8 years
1 set
2
Sealing packing
(Gas contact part)
1 year
2~4 years
1set
3
Interferometric sensor
1 year
5~8 years
1 pce
4
Circuit board
(Display)
-
5~6 years
1 pce
5
Circuit board
(MAIN)
-
7~8 years
1 pce
6
Fuse(2A)
-
8 years
2 pcs
7
Flowmeter ass’y
(Including low flow sensor)
1 year
7~8 years
2 sets
*1. The replacement interval will change depending on operating conditions and it
does not mean the guarantee period. Replacement interval may change from
inspection results.
*2. Material of sealing packing to be used for gas contact part will change according
to the kind of measuring gas.
*3. Replacement of printed circuit board is required due to the deterioration of
capacitors, etc.
*4. For assembled unit and circuit board, use pre-adjusted one.
*5. The flowmeter ass’y includes the low flow sensor.