RIGOL
User’s Guide for DSA1000 Series
6-1
6
Specification
The technical specifications are valid under the following conditions: the instrument is
within the calibration period, is stored for at least two hours at 5
℃
to 40
℃
temperature and is warmed up for 30 minutes. Unless otherwise noted, the
specifications in the manual include the measurement uncertainty.
Typical Value: characteristic performance, which 80 percent of the measurement
results will meet at room temperature (approximately 25
℃
). This data is not
warranted and does not include the measurement uncertainty.
Nominal Value: the expected average performance or a designed attribute (such as
the 50Ω connector). This data is not warranted and is measured at room temperature
(approximately 25
℃
).
Measured Value: an attribute measured during the design phase which can be
compared with the expected performance, such as the amplitude drift variation with
time. This data is not warranted and is measured at room temperature (approximately
25
℃
).
Note: Unless otherwise noted, all the data in this manual are the measurement results of multiple
instruments at room temperature. Besides, the specifications (except the TG specifications) in this
manual are for situations when the tracking generator is turned off.