Examples
:OUTP:OVP CH1,1 /*Enables the OVP function for CH1.*/
:OUTP:OVP? CH1 /*Queries the OVP on/off state for CH1. The query
returns 1.*/
4.8.13
:OUTPut:OVP:VALue
Syntax
:OUTPut:OVP:VALue
[<
source>,]{<value>|<lim>}
:OUTPut:OVP:VALue?
[<
source>][,<lim>]
:OUTPut:OVP:VALue?
[<
lim>]
Description
Sets or queries the overcurrent protection (OVP) level for the specified channel.
Parameter
Name
Type
Range
Default
<source>
Discrete
{CH1|CH2|CH3}
-
<value>
Real
Please refer to
Table 4.33 Range and default value
of overvoltage/overcurrent protection
<lim>
Discrete
{MINimum|MAXimum}
-
Remarks
•
When OVP is enabled, the output is disabled automatically if the output voltage
reaches the OVP level currently set. You can send
or
to query whether an OVP event occurred on the
specified channel.
•
If <source> is omitted, the command sets or queries the OVP level of the current
channel.
•
You can select “MINimum” to set the minimum OVP level or “MAXimum” to
set the maximum OVP level.
•
[:SOURce[<n>]]:VOLTage:PROTection[:LEVel]
to set the OVP
level for the specified channel.
Return Format
The query returns the OVP level, for example, 8.800.
Examples
:OUTP:OVP:VAL CH1,8.8 /*Sets the OVP level of CH1 to 8.8 V.*/
:OUTP:OVP:VAL? CH1 /*Queries the OVP level of CH1. The query
returns 8.800.*/
Command System