4.12.16 :TIMEr:TEMPlet:OBJect
Syntax
:TIMEr:TEMPlet:OBJect
<
obj>[,<value>|MINimum|MAXimum]
:TIMEr:TEMPlet:OBJect?
[
MINimum|MAXimum]
Description
Sets or queries the editing object of the template currently selected as well as the
corresponding current or voltage value.
Parameter
Name
Type
Range
Default
<obj>
Discrete
{V|C}
V
<value>
Real
Voltage or current range of the
channel currently selected
0
Remarks
•
Setting <obj> to “V” indicates selecting voltage to edit. <value> is used to set
the constant current value and its range is the current range of the selected
channel. At this point, you can send
and
to set the maximum and minimum values of
voltage respectively.
•
Setting <obj> to “C” indicates selecting current to edit. <value> is used to set
the constant voltage value and its range is the voltage range of the selected
channel. At this point, you can send
and
to set the maximum and minimum values of
current respectively.
Return Format
The query returns the editing object currently selected and the corresponding current
or voltage value (separated by comma). For example, the query might return V,2.000.
Wherein, “V” indicates selecting voltage to edit, and 2.000 indicates setting the
constant current value to 2 A.
Examples
:TIME:TEMP:OBJ V,2 /*Selects the editing object to voltage and sets
the constant current to 2 A.*/
:TIME:TEMP:OBJ? /*Queries the editing object and the corresponding
constant current or voltage value. The query returns V,2.000.*/
Command System
108
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.