Voltage and Current Values
During the test, the voltage and current of the output channel of the power supply
should be set to the specified values. Table 1-2 lists the rated output values and
maximum output values of the voltage and current of each channel.
Table 1-2 Voltage and current values of each channel of DP1308A
Channel
Rated Output
Voltage
Max Output
Voltage
Rated Output
Current
Max Output
Current
+6V/5A
+6 V
+6.3 V
5 A
5.25A
+25V/1A +25 V
+26.25 V
1 A
1.05A
-25V/1A -25 V
-26.25 V
1 A
1.05A
Test Result Record
Record and keep the test result of each test item. A test result record form which
provides all the test items and their corresponding performance specification limits as
well as spaces for users to record the test results, is provided in Appendix.
Tip:
RIGOL
recommends that you photocopy the test results record form before each
test and record the test results on the copies so that this form can be used
repeatedly.