RIGOL
Service Guide for DM3068
1-14
1. 6
Measurement Rate
Measurement Rate
[1]
Function
Setting
Integration Time
Readings/s 50Hz
(
60Hz
)
DC Voltage
DC Current
2-wire Resistance
4-wire Resistance
0.006 NPLC Integration Time
100(100) us
10000(10000)
0.02 NPLC
400(333) us
2500(3000)
0.06 NPLC
1.2(1) ms
833(1000)
0.2 NPLC
4(3.33) ms
250(300)
1 NPLC
20(16.7) ms
50(60)
2 NPLC
40(33.3) ms
25(30)
10 NPLC
200(167) ms
5(6)
100 NPLC
2(1.67) s
0.5(0.6)
AC Voltage
AC Current
[2]
3Hz AC Filter
0.2
20Hz
1.5
200Hz
10
200Hz
50
[3]
Frequency and
Period
[4]
1s Gate Time
1
0.1s
10
0.01s
80
0.001s
500
Capacitance
[5]
25
[1]
Auto trigger, zero trigger delay, auto zero off, auto range off, math function off and external interface off.
[2]
Use the default trigger delay setting.
[3]
The maximum rate available when trigger delay is set to 0.
[4]
20V range, fast filter, 1kHz input.
[5]
Measure 20nF capacitance on 200nF range. The measurement period changes with the capacitance under test. The maximum measurement period on
100mF is 4s (typical value).