5Hz-10Hz
0.35 + 0.08
0.35 + 0.10 0.35 + 0.10
0.035 + 0.008
10Hz-5kHz
0.15 + 0.08
0.15 + 0.10 0.15 + 0.10
0.015 + 0.008
Additional Low Frequency Errors (% of reading)
Additional Crest Factor Errors (non-sinewave)
[7]
Frequency
AC Filter
Crest Factor
Error (% of reading)
Slow
Medium
Fast
10Hz-20Hz
0
0.74
--
1 - 2
0.05
20Hz-40Hz
0
0.22
--
2 - 3
0.2
40Hz-100Hz
0
0.06
0.73
3 - 4
0.4
100Hz- 200Hz
0
0.01
0.22
4 - 5
0.5
200Hz-1kHz
0
0
0.18
>1kHz
0
0
0
[1]
Specifications are for 90-minute warm-up, slow ac filter and sinewave input.
[2]
10% overrange on all ranges except ACV 750 V and ACI 10 A ranges.
[3]
Relative to calibration standards.
[4]
Specifications are for sinewave input >5% of range. For inputs within 1% and 5% of range and <50 kHz, add 0.1% of range additional error. For
50kHz to 100kHz, add 0.13% of range additional error.
[5]
ACV 750 range limited to 8x10
7
Volt-Hz. For input over 300V rms, add 0.7mV error for each additional volt.
[6]
For continuous current > DC 7A or AC RMS 7A, 30 seconds ON and 30 seconds OFF.
[7]
For frequency blow 100 Hz, the specification of slow filter is only for sinewave input.
[8]
Specifications are for sinewave input >5% of range. For inputs within 1% to 5% of range, add 0.1% of range additional error. Specifications are typical
values for 200uA and 2mA, 2A and 10A ranges when frequency >1kHz.
Measuring Characteristics
True RMS AC Voltage
Measurement Method
AC-coupled True-RMS measurement with up to 400V DC of bias at on any range.
Crest Factor
≤ 5 at full range
Input Impedance
1MΩ ± 2% in parallel with <150pF capacitance on any range
Input Protection
750V rms on all ranges
AC Filter Bandwidth
Slow: 3Hz - 300kHz
Medium: 20Hz - 300kHz
Fast: 200Hz - 300kHz