RIGOL
Chapter 2 Front Panel Operations
2-72
DG800 User's Guide
7.
High Frequency Rejection
High frequency rejection can be used to filter out the high-frequency
components and improve the measurement accuracy in low-frequency signal
measurement. Tap the High Freq Rejection parameter selection field to
enable or disable the high frequency rejection function.
Note:
Enable the high frequency rejection when low-frequency signal with less
than 200 kHz frequency is measured to filter out the high-frequency noise
interference. Disable the high frequency rejection when high-frequency signal
with greater than 200 kHz frequency is measured. At this time, the maximum
input frequency can be 240 MHz.
Summary of Contents for DG800 Series
Page 2: ......
Page 14: ......
Page 30: ......
Page 46: ...RIGOL Chapter 2 Front Panel Operations 2 16 DG800 User s Guide Figure 2 3 Sine Waveform...
Page 57: ...Chapter 2 Front Panel Operations RIGOL DG800 User s Guide 2 27 Figure 2 5 Harmonic...
Page 136: ......
Page 142: ......