RIGOL
DG1000Z Performance Verification Guide
1-2
50Ω Load
50Ω/1W
--
Power Sensor
-30dBm to +20dBm
Agilent N8482A
Power Sensor
Connecting Cable
Used to connect the power meter
and power sensor
--
Adaptor
N (f)-BNC (m)
--
Adaptor
BNC (f)-N (m)
--
Test Result Record
Record and keep the test results of each test item. The test result record forms,
which provide all the test items and the corresponding performance specification
limits as well as spaces for users to record the test results, are provided in
“Appendix A: Test Result Record Form” of this manual.
Tip:
It is recommended that you photocopy the test result record form before each test.
During the test process, record the test results on the copies so that the forms can
be used repeatedly.
Test Notices
To achieve optimum test effect, all the test procedures should follow the following
recommendations.
1)
Make sure that the environment temperature is between 18
℃
and 28
℃
and
every test is performed under the specified operation temperature (18
℃
to
28
℃
).
2)
Before performing each test, make sure that the instrument has been warmed
up for at least 30 minutes.
3)
Before performing each test, restore the instrument to factory setting.