background image

 

5

 

 

ME12973A

 

 

 

(16)Mount the cryogenic gas suction on to the goniometer cover. Then, connect the cable to the 

goniometer cover for the cryogenic gas suction. (Fig.10) Fix the cryogenic gas suction to the 
goniometer cover without any screws as the stand of the suction is magnetic and securely 
attaches to the metal cover. 

 

(17)Turn on the switch of the main body of cryogenic gas suction and make sure that the fan is 

running inside the main body and makes a little noise.   

(18)Mount the X-ray enclosure of the 

XtaLAB mini

(19)Mount the nozzle of Low-Temperature Gas Cooler to the 

XtaLAB mini

 main body.   

(20)Remove the protective cover on the detector. Mount the collimator and beam stop.   

 

2.2    Fine alignment of the cryogenic gas suction pipe 

(1)  Turn on the switch on the main body of cryogenic gas suction. 

(2) Turn on the power of 

XtaLAB mini

 and the water chiller system.   

(3)  Launch 

CrystalClea

r and log into it. After login in the system, the dialog window 

Open 

sample

 will be displayed. Click the 

Cancel

 button. 

(4)  Click on the 

Manual Instrument Control

 under the menu bar item 

Instrument

(5)  Click the 

Initialize

 button in the

 Manual Instrument Control 

dialog window to get 

synchronization between goniometer and software. 

(6)  Move the Omega axis to -70 degrees.   

(7)  Mount the goniometer head, which holds the sample, to the goniometer.   

(8)  Launch 

RAXVIDEO

 and center the sample. 

(9)  Move the Omega axis to 90 degrees.   

Fig.10 Installation of the main body of cryogenic gas suction 

Cable to the goniometer cover 

Goniometer cover   

for cryogenic gas 

suction 

Summary of Contents for XtaLAB mini

Page 1: ...Cat No 2366A201 XtaLAB mini Cryogenic gas suction attachment Instruction manual Manual No ME12973A02...

Page 2: ...ual has to be purchased for each product If there are any missing or incorrectly collated pages in the delivered instruction manual please contact the sales representative who you purchased the instru...

Page 3: ...view 1 2 Cryogenic gas suction attachment installation and adjustment 2 2 1 Wiring the electric power supply cable 2 2 2 Fine alignment of the cryogenic gas suction pipe 5 2 3 Evaluate operation and p...

Page 4: ......

Page 5: ...cold gas flow and reduce the amount of ice accumulation Please refer to or read other manuals including XtaLAB mini system Maintenance Manual 11568C and CrystalClear Ver 2 0 0 for CCD system for Smal...

Page 6: ...alog window to get synchronization between goniometer and software 5 Move the Omega axis to 90 degrees 6 Close the CrystalClear software and turn off power to the XtaLAB mini and the water chiller sys...

Page 7: ...be 12 Confirm the cable at the rear side of goniometer goes through the hole along the Omega axis Pass the cable through the wiring duct Fig 5 near the USB ports at the rear of the XtaLAB mini 13 Remo...

Page 8: ...the suction pipe when switching on the cryogenic gas suction at the left Mount the pipe such that the height from the top to the main body is around 56 mm Fig 8 Fix the suction pipe into the main bod...

Page 9: ...and beam stop 2 2 Fine alignment of the cryogenic gas suction pipe 1 Turn on the switch on the main body of cryogenic gas suction 2 Turn on the power of XtaLAB mini and the water chiller system 3 Lau...

Page 10: ...according to the following conditions for one image within flow bar step Initial Images Start Angle End Angle Width Exposure time sec 30 31 1 0 16 3 Inspect the image and make sure there is not a sha...

Page 11: ...tion Rigaku Americas Corporation Rigaku Europe SE Head Office 3 9 12 Matsubara cho Akishima shi Tokyo 196 8666 Japan Phone 81 42 545 8189 Fax 81 42 544 9223 e mail rinttyo rigaku co jp 9009 New Trails...

Reviews: