(14)Connect the adapter cable to the power outlet.   

(15)Mount the suction pipe into the cryogenic gas suction. Refer to the Fig.7, the cooling gas 

alignment guide must be positioned at the bottom of the suction pipe when switching on the 
cryogenic gas suction at the left. 

 

 

Mount the pipe such that the height from the top to the main body is around 56 mm. (Fig.8) 

 

 

Fix the suction pipe into the main body using the slotted set screw. (Fig.9) 

 

 

56 mm 

Fig.8 Location of the suction pipe 

Cooling gas alignment guide 

 

Fig.7 Installation of suction pipe 

 

Set screw to fix the suction pipe 

Fig.9 Fixing the suction pipe 

Summary of Contents for XtaLAB mini

Page 1: ...Cat No 2366A201 XtaLAB mini Cryogenic gas suction attachment Instruction manual Manual No ME12973A02...

Page 2: ...ual has to be purchased for each product If there are any missing or incorrectly collated pages in the delivered instruction manual please contact the sales representative who you purchased the instru...

Page 3: ...view 1 2 Cryogenic gas suction attachment installation and adjustment 2 2 1 Wiring the electric power supply cable 2 2 2 Fine alignment of the cryogenic gas suction pipe 5 2 3 Evaluate operation and p...

Page 4: ......

Page 5: ...cold gas flow and reduce the amount of ice accumulation Please refer to or read other manuals including XtaLAB mini system Maintenance Manual 11568C and CrystalClear Ver 2 0 0 for CCD system for Smal...

Page 6: ...alog window to get synchronization between goniometer and software 5 Move the Omega axis to 90 degrees 6 Close the CrystalClear software and turn off power to the XtaLAB mini and the water chiller sys...

Page 7: ...be 12 Confirm the cable at the rear side of goniometer goes through the hole along the Omega axis Pass the cable through the wiring duct Fig 5 near the USB ports at the rear of the XtaLAB mini 13 Remo...

Page 8: ...the suction pipe when switching on the cryogenic gas suction at the left Mount the pipe such that the height from the top to the main body is around 56 mm Fig 8 Fix the suction pipe into the main bod...

Page 9: ...and beam stop 2 2 Fine alignment of the cryogenic gas suction pipe 1 Turn on the switch on the main body of cryogenic gas suction 2 Turn on the power of XtaLAB mini and the water chiller system 3 Lau...

Page 10: ...according to the following conditions for one image within flow bar step Initial Images Start Angle End Angle Width Exposure time sec 30 31 1 0 16 3 Inspect the image and make sure there is not a sha...

Page 11: ...tion Rigaku Americas Corporation Rigaku Europe SE Head Office 3 9 12 Matsubara cho Akishima shi Tokyo 196 8666 Japan Phone 81 42 545 8189 Fax 81 42 544 9223 e mail rinttyo rigaku co jp 9009 New Trails...

Reviews: