9
To prevent damage to the
battery, charge it when
the light turns yellow. The indicator light
will extinguish 10 minutes after the last
activation of the up/down switch. It will
turn on again if the up/down switch is used
or if a battery is inserted into the battery
box.
If the battery level becomes critically low,
indicated by the battery level light turning
red, the microprocessor will disable the
up/down switch. However, the emergency
lowering button will continue to function.
The emergency lowering button will continue
to function even if the battery level has
dropped below the threshold for operation
of the up/down switch. It overrides all other
inputs.
To prevent injury, all
caregivers should
become familiar with the location and
operation of the emergency stop button
and the emergency lowering button
before using the E-Pacer.
Emergency stop button
To stop the E-Pacer motor in the event
of an emergency, press the round red
emergency stop button located on the front
of the battery box (see Figure 10a). Twist it
clockwise to reset it.
Emergency lowering button
The emergency lowering button (A),
located on the front of the battery box,
provides a means of lowering the body
support if the up/down switch fails.
Press the emergency lowering button to
lower the body support as required (see
Figure 10b).
NOTICE
CAUTION
Figure 10a
Figure 10b
A
A