13
Scale
The scale has two purposes: it can be
used to measure a client’s weight, and
to measure the weight a client is bearing
during ambulation.
To use the scale, turn on the display and
use the lb/kg button to display either
pounds or kilograms. Before approaching
the client who will be weighed, attach
all the components you will need for this
particular client, (leg straps, arm supports,
etc.) and then press the ZERO button to
zero the scale. Proceed to lift the client.
Once a client is completely supported by
the E-Pacer, the client’s weight will be
displayed (see Figure 13a).
There are two ways to measure
weight-bearing during ambulation:
Method 1:
Turn on and zero the scale
as previously described. Once the client
is lifted and in the walking position, the
weight displayed is the amount of weight
that the E-Pacer is bearing.
Method 2:
Using the thigh supports and
the seated transfer procedure, lift the client
with the E-Pacer. With the client’s full
weight displayed on the scale, press the
zero button. Next, re-position the client
for ambulation. The weight the client is
bearing will be displayed as a negative
number.
To replace the scale batteries,
push
outward on the small tab at the rear of the
scale housing and lift the battery cover (a
small, fl at head screwdriver inserted into
the slot at the back of the lid may be used
to gently pry it up). Insert fresh AA Alkaline
batteries; the cells must be oriented as
shown by the inscription in each battery
receptacle (see Figure 13b).
Figure 13a