7. Position the load cell to maintain the dimension.
8. Route the load cell cables near each corner to ensure the cable is free from possible contact with each foot.
9. Do not cut excess load cell cables.
10. Make a strain relief loop before routing the cables through the conduit.
11. Coil extra cable before it enters the junction box.
12. Pass the end of the load cell cable through the cable fittings in the NEMA Type 4X junction box.
13. Corner correction, trimming, and calibration are necessary after replacing a load cell. Follow the instructions in
and
4.3
Troubleshooting Guide
Table 4-1. Troubleshooting Guide.
Symptom
Probable Cause
Solution
System does not operate--no
display
Power disconnected
Check and reconnect
Indicator fuse blown
Replace fuse, check for cause
Interface cable cut or disconnected
Repair
Signal leads incorrectly installed at indicator
Install according to indicator installation manual
Display stays at zero
Faulty load cell connections
Check cable connections in J-box at indicator
Faulty indicator
Service the indicator
Erratic weights
Vibration near scale
Remove source of vibration or move scale
Platform not level to within 1/4”
Level the platform by adjusting feet or shimming
Load cell or cable water damage
Replace load cell or cable
Debris under load cells or platform
Clean
Faulty indicator
Use simulator to test indicator stability; service indicator
Consistently high or low
weights
Indicator not properly adjusted to zero
Zero the indicator according to operation manual
Platform binding
Adjust clearance for free platform movement
Indicator not calibrated
Calibrate according to indicator’s manual and
Faulty load cells
Test and, if necessary, replace load cells
Feet touching deck’s underside
Adjust feet downward to provide clearance