RoughDeck QC-X Floor Scale With Gas Springs
12
Visit our website
3.3
Calibration
See the indicator operation manual to determine the indicator calibration procedures.
It is recommended that the scale be exercised before calibration ensure everything is seated.
1. Load the scale to near capacity two or three times.
2. Ensure there is no load on the scale and place the indicator in the calibration mode.
3. Perform a zero calibration.
4. Place test weights equal to 70% – 80% of the scale capacity on the platform. When using several weights, distribute
them evenly around the platform.
5. Perform a span calibration.
6. Remove the test weights and check the zero reading.
7. Repeat the calibration process, if necessary.