Appendix
© Rice Lake Weighing Systems
●
All Rights Reserved
63
11.3 Continuous Data (Stream) Output Formats
When the trigger setting for a port is set to STRIND or STRLFT, data is continuously streamed from the appropriate port in one
of the four fixed format options or a custom format option.
Fixed Format Options:
• Rice Lake Weighing Systems (
• Cardinal (
)
• Avery Weigh-Tronix (
)
• Mettler Toledo (
11.3.1 Rice Lake Weighing Systems Stream Format (
RLWS
)
Figure 11-1. Rice Lake Weighing Systems Stream Data Format
11.3.2 Cardinal Stream Format (
cardnal
)
Figure 11-2. Cardinal Stream Data Format
<STX> <POL> <wwwwwww> <UNIT> <G/N> <S> <TERM>
ASCII 02
decimal
Polarity:
<space> = Positive
<–> = Negative
Weight: 7 digits, right-justified, dummy
zeroes, decimal point with no leading
zeroes except for leading zero immediately
preceding the decimal point. Leading
zeroes transmitted as spaces.
L = pounds
K = kilograms
T = tons
G = grams
O = ounces
G = Gross
N = Net
Status:
<space> = valid
I = Invalid
M = Motion
O = Over/under range
Z = COZ
<CR> <LF>
or <CR>
Carriage
Return
Polarity:
<+> for positive
"–" for negative
Six digits, right justified, floating decimal
(follows weight if not embeded),
leading zeroes are present.
g for Gross
n for Net
Hex value of <03>
is entered in string
lb = pound
kg = kilogram
g = gram
tn = ton (short)
t = ton (metric)
oz = ounce
<space> = none
Space
Character
<CR> <POL> <wwwwww> <S> <SP> <UNIT> <SP> <G/N> <SP> <SP> <ETX>
Space character
(2
)
Space
Character
Status:
<space> = valid or invalid
m = motion
o = over/under range