Ranger Simulators 1, 3 and 5 Operation
3
2.0
Ranger Simulators 1, 3 and 5 Operation
To get the most out of the simulator, remember the following:
• Dampness, water, or other contaminants on the face of the simulator can cause leakage current between
terminals. Keep the face and the terminals clean and dry when using the Ranger Series simulator.
• The binding posts are heavily plated, but should some corrosion occur, clean the posts before attempting to
use the unit.
• If the simulator has been at an unusual temperature environment for many hours (i.e., left in a truck during
winter or summer), allow several minutes for the unit to stabilize to the ambient temperature of the scale
before using.
• Excitation voltage should not exceed 15 volts DC or AC (RMS). Higher voltages will degrade the accuracy
and possibly damage your simulator.
• When using the Ranger 5 simulator for testing stability or drift, or when recalibrating an indicator, make sure
the vernier range select switch is in the Off position.
1
2
3
4
5
6
7
Note
The vernier is included as a diagnostic and setup tool only.
Figure 2-1. Ranger 5 Control Location
1. Vernier Range Selector
OFF: Rotary
Selector
value
FINE:
Rotary Selector value -0.02 mV/V to +0.5 mV/V
MEDIUM: Rotary Selector value -0.05 mV/V to +1.5 mV/V
COARSE: Rotary Selector value -0.12 mV/V to +3.0 mV/V
2. Locking Vernier Dial: 10-turn adjustment of selected ranges listed above
3. Rotary Selector: Fixed calibration steps of 0.5 mV/V from -0.5 to 3.0 mV/V + 1.5mV/V step to 4.5 mV/V
4. +Excitation Input terminal post
5. +Signal Output terminal post
6. -Signal Output terminal post
7. -Excitation Output terminal post