Ranger Simulator Operation Guide

3.0

Ranger 7 Simulator Operation

0.5mV/V

 ZERO

0.01mV/V

0.1mV/V

+EXC

–EXC

+SIG

–SIG

X4

X1

X.5

.04
.03
.02
.01
.00 

.06
.07
.08
.09
.10 

.05

.4
.3
.2
.1 

.0 

 2.5 

2.0
1.5
1.0
0.5 

0.0 

+    – 

EXC

.0 

PUSH TO ENABLE

HOLD TO CHANGE RANGE

0.5mV/V

VARIABLE

0.1mV/V

PUSH TO ENABLE

±25μV/V

OFFSET

ON

– 

High Resolution Load Cell Simulator

Switch One

Switch Two

Switch Three

Variable Switch

Offset Switch

EXC LEDs

Toggle Switch

Binding Posts

Figure 3-1. Ranger 7 Simulator

3.1

Standard Operation

3.1.1

Switches

The three switches across the bottom front face of the Ranger 7 work together up to 3 mV/V. 

• Switch one maximum = 2.5
• Switch two maximum =   .4 (has two .0 settings, they are identical settings)
• Switch three maximum = .10 (

ZERO

 position puts the mV/V output setting to zero regardless of the other 

switch settings)

Note

When switch three is set to ZERO, only the offset switch can vary the output in this setting. This allows an easy 
way to check AZM settings as well as quick return to zero for time testing. 

Example: 

Switch one is set to 2mV/V, switch two is set to 0.0 and switch three is set to .00. By setting switch three to the 
ZERO position, the output will instantly jump between 2mV/V and 0mV/V.

3.1.2

Toggle Switch Load Control

X1

 mode – load is 350

, equivalent to a single standard load cell

X.5

 mode – load is 700

, providing a lightly loaded system

X4 

mode – load is 87.5

 which is equivalent to a 4 load cell platform scale

3.1.3

EXC LEDs

An indication of excitation voltage presence, indicating the polarity of the excitation to aid in proper wiring. 
If both LEDs are lit, it indicates AC excitation. 

Note

The Ranger 7 works with all polarities of excitation as long as the peak or DC voltage is between 3.5V and 12V.

3.1.4

To Simulate a Larger Load

1. Place a power resistor between the EXC+ and EXC- binding posts. Use a 22

 7.5W resistor for excitation 

up to 10V. 

2. Put the load switch in the 

X.5

 position. Calculate the required resistance using 

R=(700 x X)/(700 - X)

X

 is the 

desired load. 

Summary of Contents for Ranger 1

Page 1: ...105354 Rev A Ranger Series Load Cell Simulators Operation Manual...

Page 2: ......

Page 3: ...riable Setting 7 4 0 Troubleshooting With a Simulator 8 5 0 Product Specifications 9 Limited Warranty 10 Rice Lake Weighing Systems All rights reserved Printed in the United States of America Specific...

Page 4: ...ii Ranger Series Operation Manual Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Page 5: ...re 1 1 Ranger 1 3 and 5 Simulators Features Ranger 1 Bridge impedance of 700 ohms 10 turn continuously variable output from 0 to 3 mV V Ranger 3 Bridge impedance of 350 ohms Nine calibrated outputs ei...

Page 6: ...or 2 Ranger Series Operation Guide Figure 1 2 Ranger 7 Simulator Ranger 7 Load Impedance Three settings by toggle switch X0 5 700 X1 350 1 standard Load Cell X4 87 5 4 Load Cell Platform Output Range...

Page 7: ...hould not exceed 15 volts DC or AC RMS Higher voltages will degrade the accuracy and possibly damage your simulator When using the Ranger 5 simulator for testing stability or drift or when recalibrati...

Page 8: ...oltage value from step 2 Example 2 99825 x 9 9875 29 945 mV This is the full scale millivolt output of the scale at the total load cell capacity individual load cell capacity x the number of load cell...

Page 9: ...librated Pre Calibrating an Indicator Using a Simulator Ranger 3 and 5 Only For this example we will pre calibrate an indicator that will be connected to a single load cell application The cell capaci...

Page 10: ...n this setting This allows an easy way to check AZM settings as well as quick return to zero for time testing Example Switch one is set to 2mV V switch two is set to 0 0 and switch three is set to 00...

Page 11: ...function 3 3 Variable Press the Variable switch to enable disable the function There are two ranges of variable mV V 3 3 1 Coarse Control Range The range is 0 to 500 V V 0 to 0 5mV V plus the setting...

Page 12: ...s substituted for the load cells and the value displayed for the zero and one of the other calibrated settings i e 1 mV V are recorded later recalibration or repair can be greatly simplified by direct...

Page 13: ...equal to the 0 5mV V switch plus 0 0 5mV V Range 2 is equal to the 0 5mV V switch plus the 0 1mV V switch plus 0 0 1mV V Enabled by pressing the switch Switchable Offset Off or switch setting 25 V V...

Page 14: ...caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modified...

Page 15: ......

Page 16: ...8 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 88 2349171 www ricelake com www ricelake mx www ricelake eu www ricelake co in m ricelake com Rice Lake Weighin...

Reviews: