10.0 Appendix
10.1 Two-Channel IQube Diagnostics
The two-channel iQube provides a number of diagnostic tests. See the table below for a list of diagnostic error
messages. It also describes the tests that can be enabled or disabled in the diagnostics of the two-channel iQube.
Table 10-1. Two-Channel iQube Diagnostic Settings
Test
Definitiion
Description
920i Default
Revolution III/
two-channel Default
Zero
Reference
Which cell is out of tolerance
at zero
Threshold: the upper limit of the zero reading
of the cells
0.125mV
0.200 mV
Percent threshold: the percentage of zero
reading of adjacent cells
25%
20%
Delay: time to wait until the next testing
process
900 seconds
900 seconds
Load Cell
Drift
Checks for load cell outputs
that have changed over time
with load applied
Limit
0.025mV
0.025mV
Percent limit
10%
10%
Time
30 seconds
30 seconds
Load: minimum load needed to initiate the
test
1000 lb
1000 lb
Peak to
Peak Limit
Check for noisy load cell
output not attributed to
environmental conditions
Limit
0.125mV
0.250mv
10.1.1 Diagnostic Setup
Parameters associated with each of the two-channel iQube’s diagnostics test can be set using the 920i menus, iRev,
or Revolution III.
When using the two-channel iQube with the 920i indicator, a Diagnostics softkey can be configured to display
diagnostic information when in weigh mode. Use the FEATURE menu, SOFTKEYS submenu, to select the
diagnostics softkey.