MSI-3460 Challenger 3
22
Visit our website
5.7 Auto Zero Maintenance (AZM)
An auto zeroing maintenance mechanism is used to adjust the zero reading to the center-of-zero. The center-of-zero is defined
as the weight reading within 1/4 d of zero. AZM continuously adjusts zero to maintain the center-of-zero. It is recommended that
AZM is on to maintain the highest accuracy.
There are circumstances when it should be turned off. This can happen when minor variations of weight occur while picking up
scale attachments and the variations fall within the AZM capture window. The AZM capture window (usually 1 d) and capture
time (usually 8 seconds) can be adjusted by Rice Lake Weighing Systems to meet custom requirements.
The settings of AZM are dictated in Legal for Trade standards and cannot be adjusted.
Use the following steps to set up a Legal for Trade standard settings.
1. Enter
Configuration
mode (
).
displays.
2. Press
to scroll to
.
3. Press
.
displays.
4. Press
to scroll to
.
5. Press
.
The current setting is displayed.
6. Press
to toggle between
or
.
7. Press
.
displays.
8. Press
twice to save settings.
displays briefly and exits setup.
5.8 Zero Power Up (0.P-UP)
This feature will cause the unit to automatically zero after the unit is turned on. Default is
.
1. Enter
Configuration
mode (
).
displays.
2. Press
to scroll to
.
3. Press
.
displays.
4. Press
to scroll to
.
5. Press
.
The current setting is displayed.
6. Press
to toggle between
or
.
7. When desired value is displayed, press
.
displays.
8. Press
twice to save settings.
displays briefly and exits setup.
USER
F
TARE
USER
F
TARE
USER
F
TARE
ZERO
USER
F
TARE
USER
F
TARE
USER
F
TARE
ZERO