Calibration
© Rice Lake Weighing Systems
●
All Rights Reserved
21
5. Press
.
A flashing
displays.
6. Press
to enter the C-Cal value. The far left digit will flash indicating that number should be entered.
7. Press
to scroll through the numbers and
to enter each digit.
8. When correct number is displayed, press
to save. The display will read
followed by
’
.
9. Press
to exit
C-Cal Setup
menu.
10. Press
to store the calibration and return to the weigh mode.
displays briefly.
5.6 Standard Menu
Table 5-1. Standard Menu Selections
Use the following steps to set up standard settings.
1. Enter
Configuration
mode (
).
displays.
2. Press
. to scroll to
.
3. Press
.
displays.
4. Press
.
The current standard setting is displayed.
5. Press
to scroll to the desired standard.
6. Press
to set the standard.
7. Press
.
is displayed.
8. Press
twice to exit setup and store all changes.
is displayed briefly, then the unit goes to
Weigh
mode.
Selection
Description
Industrial (
)
This is the most common setting for the MSI-3460; With the Industrial standard, there is full range zero, access to units switching,
filters, and peak hold
Handbook 44 (
) Enables only approved features per the NTEP HB-44 rules and regulations; Access is denied to Peak Hold, and the zero range
may be limited; The Filter menu is moved to the calibration setup menu, so filters are only accessible through the calibration seal
R-76 (
)
Sets the scale to enable only approved features per OIML R-76; Only kg weight units are available; The zero range is limited to
4% (-1 to +3% relative to calibrate zero); Net/Gross function is temporary; Once net weight is established, pushing an F key set for
Net/Gross will cause a maximum 5 second display of the gross weight; Clear the Tare to display gross weight constantly;
Other metrological aspects are changed to meet R-76 requirements; Only stable weights may be printed; Negative weight display
is limited to -20d
1Unit (
)
The 1unit standard is exactly the same as Industrial, except units switching is inhibited; This is useful for metric only countries;
Another use of the 1Unit standard is to allow the scale to be calibrated in units other than lb or kg, since conversions are
eliminated; Contact Rice Lake for more information on the standards settings
TARE
USER
F
USER
F
TARE
TARE
ZERO
ZERO
USER
F
TARE
TARE
USER
F
TARE
ZERO
ZERO