Configuration
25
PROGRM menu parameters continued
DSPHLD
DSPMOD
HDELTA
SMPTIM
Specifies whether operator wants to use the display hold mode, set the hold delta percentage
and the display hold sampling time. See sub-menu parameter descriptions below.
DSPHLD sub-menu parameters
DSPMOD
DSABLE
ENABLE
Display hold mode – Specify whether display hold mode is to be used or not. If it is enabled,
the indicator will sample the scale for SMPTIM and display the average weight.
HDELTA
80
10-99
Delta percentage that the display value must drop before scale returns to live weighing. Valid
values are 10-99 percent with a default of 80 percent.
SMPTIM
4
1-10
The amount of time in seconds that the indicator will sample the scale before it displays the
average of those samples. Valid values are 1-10 seconds with a default of 4 seconds.
PROGRM menu parameters continued
REGULA
NTEP
CANADA
NONE
OIML
Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.
• OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than
zero. NONE allows tares to be acquired at any weight value.
• OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at
no load. NONE allows tares to be cleared at any weight value.
• NTEP and OIML modes allow a new tare to be acquired even if a tare is already present.
In CANADA mode, the previous tare must be cleared before a new tare can be acquired.
• NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net
mode as long as the current weight is within the specified ZRANGE. In OIML mode, the
scale must be in gross mode before it can be zeroed; pressing the ZERO key in net mode
clears the tare.
CONSNU
000000
number
Consecutive numbering. Allows sequential numbering for print operations. The consecutive
number value is incremented following each print operation.
The initial value of this parameter is set to the start up value specified on the CONSTU
parameter. Changing either CONSTU or CONSNU immediately resets the consecutive number
used for printing.
CONSTU
000000
number
Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value
used when the indicator is powered on.
UID
1
number
Unit ID. Specify a unit identifier for the indicator in the range 1–999999. The unit ID can be
added to print ticket formats to identify the indicator used to generate the ticket.
PROGRM Menu
Parameter
Choices
Description
Table 3-4. Program Menu Parameters (Continued)