250-10-2 Bariatric Handrail Scale

14

Visit our website 

www.RiceLake.com

4.5 Default Menu

The default menu is used to return the scale back to its factory settings and is shown below.

Figure 4-4. Default Menu

Use the following steps to return the settings back to their factory default.

1. Press 

. The display shows a default value of 

NO.

2. To change to 

YES

, use 

 and 

 to adjust the value.

3. Press 

 and the display shows 

DONE

.

4. Press 

 to return to top level level menu.

Selecting YES and pressing 

 will reset to factory defaults settings without changing the calibration and will 

return you to Weigh mode. 

4.6 Scale Calibration

Use the following steps to calibrate the scale.

1. Press 

 and 

 simultaneously to power on the scale.

2. The unit cycles through its startup function and continues to display the software version. Continue to hold both keys 

until 

Id

 appears.

3. Access the setup switch located in the back of the scale to enter the setup parameters for the scale. Use a small paper 

clip, small screwdriver or other similar object to press the setup switch on the back of the indicator (under the cover).

4.

Prog

 displays. Press 

 to toggle along the parameter menu.

5.

Calib

 displays and enter the calibration parameters.

6. Press 

 and a numeric value is displayed which represents the amount of weight that is used for calibration. 

Lb will be flashing. To switch between lb and kg, press 

 or 

. Once a unit is selected, press 

 

and the 

right most digit will be flashing.

7. To change the calibration load value, press 

 or 

 to increment/decrement the flashing digit.

8. Use 

 to move the flashing digit to the left or right.

9. Once all the digits have been entered, press 

 and 

Clear

 displays.

10. Make sure the scale platform is clear of weight and press 

 again then 

=====

 displays.

11. A request to put the chosen load on the platform is displayed by 

Put xxx.xx

.

12. Put the chosen weight on the platform and press 

=====

 displays then 

Save

 displays.

DEF

CALIB

PROG

DONE

  NO

   YES

 TEST

ENTER

ENTER

BMI

Note

ENTER

ENTER

BMI

ENTER

ENTER

BMI

ENTER

ENTER

ENTER

Summary of Contents for 250-10-2

Page 1: ...250 10 2 Bariatric Handrail Scale Software Version 11525 Technical Manual PN 151436 Rev D August 17 2022...

Page 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Page 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...

Page 4: ...ghing 6 3 3 Hold Release Function 6 3 4 Preset Tare 6 3 5 Toggle Tare 6 3 6 Using the Body Mass Index BMI Function 7 3 6 1 LB Mode 7 3 6 2 KG Mode 7 3 7 Troubleshooting 8 4 0 Configuration 9 4 1 Setup...

Page 5: ...se limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio freq...

Page 6: ...rt the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any...

Page 7: ...e must be returned or moved The product must be properly packed with sufficient packing materials Whenever possible use the original carton and packing materials when shipping the scale back Damage ca...

Page 8: ...age of 25 hours of continuous use To install the batteries 1 Turn thumbscrew counterclockwise then remove battery cover 2 Insert batteries into the battery chamber as illustrated Figure 2 2 Battery Ch...

Page 9: ...sing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks for the...

Page 10: ...used by the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annu...

Page 11: ...the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that the scal...

Page 12: ...D too high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left...

Page 13: ...enus and or parameters Press to enter a displayed menu and or parameter Press or to scroll through values Press to save the displayed selection and move to the next parameter 4 2 2 Enter Numbers Press...

Page 14: ...h located in the back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch 4 Once the setup switch is pre...

Page 15: ...the Up Down arrow keys to increment decrement numbers 7 When done press ENTER key to move to the next parameter LOAD LOAD This is the amount of weight applied during calibration Can also be changed in...

Page 16: ...2 5 10 20 50 100 Values in lb 1 2 5 10 20 50 100 200 0 2 0 5 1 0 2 0 5 0 10 0 20 0 0 1 The display toggles between a numeric value and ROUND The decimal point location is set to the DISP parameter dis...

Page 17: ...the following steps 1 Press the ENTER key to change values 2 Use the Up Down arrow keys to change the available parameters 3 When done press ENTER key to move to the next parameter UNITS UNITS Units...

Page 18: ...f the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch on the back of the indicator under the cover 4 Prog d...

Page 19: ...the scale until ID flashes 2 Press again 3 Continue to press to scroll through the various menu items 4 Once complete press again then Done displays 5 Press to start the weighing process Figure 4 5 T...

Page 20: ...the weight was either in motion or out of range If displaying weight and not BMI the scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Example 60 1 KG PATIENT SP...

Page 21: ...rotocol Commands Name ESC Character ESC Value with Parameters Description Reading R R Tells PC the scale is sending a reading immediately following this is the value that is sent Example ESC R ESC W02...

Page 22: ...AD Z Zero the scale F Show flash values used for the first flash process L USB On Off not available on USB communication Table 5 5 Maintenance Protocol Commands Escape Protocol Example 1 Escape Protoc...

Page 23: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a P...

Page 24: ...personnel Go through the following steps for basic maintenance Check the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for...

Page 25: ...off can be configured Data Communications RS 232 with RJ 45 jack USB connection Selectable baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature 50 F t...

Page 26: ...250 10 2 Bariatric Handrail Scale 22 Visit our website www RiceLake com...

Page 27: ......

Page 28: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...

Reviews: