background image

250-10-2 Bariatric Handrail Scale

8

Visit our website 

www.RiceLake.com

3.7 Troubleshooting

Refer to the following table to check and correct any failure before contacting service personnel.

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead batteries

Replace batteries or connect to AC power

Faulty electrical outlet

Use a different electrical outlet

Bad power supply

Replace adapter

Questionable weight or the scale does not 

zero

External object is interfering with the scale Remove the interfering object from the scale
Display did not show 0.0 before weighing Help the patient off the scale, zero the scale and begin the 

weighing process again

Scale is not placed on a level floor

Ensure scale is level and begin the weighing process again

Scale is out of calibration

Check the weight with a certified calibration weight 

Scale base is touching floor during a 

weighment

Adjust height of feet so fingers can slide between the base 

of scale and the floor all the way around the platform

The display shows a 

STOP

 message

The load on the scale exceeds the 

capacity of the scale

Remove the excess weight and use the scale according to 

manufacture specifications

The display shows 

LO Bat

 message

The battery is low

Replace batteries

The display shows E and Err messages as detailed below
E06

Identifier - ADC

AD too high

E07

AD too low

E10

Overload

Scale has been overloaded. Remove load from scale

E4L

BAT

Battery low, but still usable- one bar left on indicator display

E4U

Battery low and unstable - no bars left on indicator display

E11

CAL

Calibration Error - recalibrate scale

Err 1

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Err 2

Low saturation state (low A/D)

The load cell is not connected properly; Check the cables 

and mechanical connections; if the problem persists, replace 

the set of load cells

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight; Cannot calibrate

Check the load cell mechanical surroundings and ensure 

nothing is contacting the load cell and that the cables are 

properly welded

Err 7

Scale isn’t moving

Make sure feet are installed on the scale. Turn the feet all 

the way in and then back them out three full turns, then level 

the scale

SAT

Damaged load cell cable

Replace load cell cable

Load cell cable may be plugged into wrong 

connection port

Ensure cable is connected to the load cell connection port.

Note: 

Load cell connection point is located underneath the 

curved plastic cover of the indicator. Remove four back 

retaining screws, remove curved back cover to access load 

cell connection point.

Table 3-2.   Troubleshooting Table

Summary of Contents for 250-10-2

Page 1: ...250 10 2 Bariatric Handrail Scale Software Version 11525 Technical Manual PN 151436 Rev D August 17 2022...

Page 2: ...respective companies All information contained within this publication is to the best of our knowledge complete and accurate at the time of publication Rice Lake Weighing Systems reserves the right to...

Page 3: ...ing or obtained by calling 715 234 9171 and asking for the training department Revision History This section tracks and describes the current and previous manual revisions for awareness of major updat...

Page 4: ...ghing 6 3 3 Hold Release Function 6 3 4 Preset Tare 6 3 5 Toggle Tare 6 3 6 Using the Body Mass Index BMI Function 7 3 6 1 LB Mode 7 3 6 2 KG Mode 7 3 7 Troubleshooting 8 4 0 Configuration 9 4 1 Setup...

Page 5: ...se limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio freq...

Page 6: ...rt the scale while someone is on the scale Do not allow minors children or inexperienced persons to operate this scale Do not use in the presence of flammable materials Do not use this product if any...

Page 7: ...e must be returned or moved The product must be properly packed with sufficient packing materials Whenever possible use the original carton and packing materials when shipping the scale back Damage ca...

Page 8: ...age of 25 hours of continuous use To install the batteries 1 Turn thumbscrew counterclockwise then remove battery cover 2 Insert batteries into the battery chamber as illustrated Figure 2 2 Battery Ch...

Page 9: ...sing the BMI key enables access to the BMI Body Mass Index mode defaults when scale is turned on The patient is gets on the scale weight stabilizes and press the BMI key The display then asks for the...

Page 10: ...used by the patient 1 Press to turn on the scale 0 0 appears on the display along with the ZERO annunciator 2 Place additional item s on the scale 3 Press until the display returns to 0 0 and NET annu...

Page 11: ...the height value 5 Press 6 The BMI value and BMI annunciator are shown on the display Press to return to the Weighing mode and the BMI function will be turned off 3 6 2 KG Mode 1 Ensure that the scal...

Page 12: ...D too high E07 AD too low E10 Overload Scale has been overloaded Remove load from scale E4L BAT Battery low but still usable one bar left on indicator display E4U Battery low and unstable no bars left...

Page 13: ...enus and or parameters Press to enter a displayed menu and or parameter Press or to scroll through values Press to save the displayed selection and move to the next parameter 4 2 2 Enter Numbers Press...

Page 14: ...h located in the back of the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch 4 Once the setup switch is pre...

Page 15: ...the Up Down arrow keys to increment decrement numbers 7 When done press ENTER key to move to the next parameter LOAD LOAD This is the amount of weight applied during calibration Can also be changed in...

Page 16: ...2 5 10 20 50 100 Values in lb 1 2 5 10 20 50 100 200 0 2 0 5 1 0 2 0 5 0 10 0 20 0 0 1 The display toggles between a numeric value and ROUND The decimal point location is set to the DISP parameter dis...

Page 17: ...the following steps 1 Press the ENTER key to change values 2 Use the Up Down arrow keys to change the available parameters 3 When done press ENTER key to move to the next parameter UNITS UNITS Units...

Page 18: ...f the scale to enter the setup parameters for the scale Use a small paper clip small screwdriver or other similar object to press the setup switch on the back of the indicator under the cover 4 Prog d...

Page 19: ...the scale until ID flashes 2 Press again 3 Continue to press to scroll through the various menu items 4 Once complete press again then Done displays 5 Press to start the weighing process Figure 4 5 T...

Page 20: ...the weight was either in motion or out of range If displaying weight and not BMI the scale will send out the following 21 character string xxxxxxxxx SP uu SP mmmmm SP CR LF Example 60 1 KG PATIENT SP...

Page 21: ...rotocol Commands Name ESC Character ESC Value with Parameters Description Reading R R Tells PC the scale is sending a reading immediately following this is the value that is sent Example ESC R ESC W02...

Page 22: ...AD Z Zero the scale F Show flash values used for the first flash process L USB On Off not available on USB communication Table 5 5 Maintenance Protocol Commands Escape Protocol Example 1 Escape Protoc...

Page 23: ...the use of the USB driver Driver protections may need to be temporarily disabled on Windows 10 or later computers to allow for the installation of the USB driver 1 Connect the scale s indicator to a P...

Page 24: ...personnel Go through the following steps for basic maintenance Check the overall appearance of the entire scale for any obvious signs of damage Inspect the condition of the AC power adapter cord for...

Page 25: ...off can be configured Data Communications RS 232 with RJ 45 jack USB connection Selectable baud rate default 9600 8 bits No parity 1 stop bit No handshaking Environmental Operating Temperature 50 F t...

Page 26: ...250 10 2 Bariatric Handrail Scale 22 Visit our website www RiceLake com...

Page 27: ......

Page 28: ...54868 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 Rice Lake Weighing Systems Specifications subject to change without notice www ricelake com PN...

Reviews: