Um = 30 VDC
Um = 125 / 250 VAC
Wiring diagramm RHE 08 standard
E08W-E_v5
common connection ( 0 Volt )
for frequency / pulses and
digital outputs 1 to 3
digital output 3 ( passive )
function programmable
digital output 2 ( passive )
function programmable
digital output 1 ( passive )
function programmable
digital output ( passive )
for mass flow rate
frequency / pulses
external voltage
max. 24 VDC
analog output 1 ( active )
current loop : 0/4 - 20 mA
max. 470 Ohm
analog output 2 ( active )
current loop : 0/4 - 20 mA
max. 470 Ohm
digital input 2 ( passive )
function programmable
external voltage
max. 24 VDC
digital input 1 ( passive )
function programmable
15.04.2004
- or -
1
2
3
4
5
6
7
8
9
10
11
-
+
17
18
19
20
21
22
23
15
16
-
+
-
+
RHE 08
12
13
14
24 VDC
115/ 230 VAC
-
+
PE
29
28
30
N
L1
26
27
25
-
+
24
2,7K
2,7K
1
3
9
8
7
6
5
4
2
max. 200 m
(Lc = 0,9 mH/km, Cc = 0,2 uF/km)
1 / 1
H.G.Rudolph
U.Hettrich
Draw. - Rev.
Sheet
Project
Appr.
Drawn
Date
created :
Customer
External intrinsically safe
zeroing contact (option), is
subject of a separate EEx - certificate.
Note :
EEx - version : In - and outputs
(connectors 15 to 27) are galvanically
isolated. Sensor connections (1 to 9) are
intrinsically safe.
HT- SENSORS (High Temperature):
Screen to ground connection MUST
BE done. An additional potential
equalising cable is required (see Manual).
NT/ ETx - Sensors:
please DO NOT connect !
Not EEx - version : In - and outputs
(connectors 15 to 27) are galvanically
isolated. Sensor connections (1 to 9) are
not intrinsically safe.
Terminals Um Im
1 - 2 8,6 V 141 mA
3 - 4 7,4 V 29 mA
3 - 5 7,4 V 29 mA
6 - 7 7,4 V 29 mA
8 - 9 7,4 V 29 mA
12 - 13 7,4 V 29 mA
14 - 16 14 V 76 mA
15 - 16 14 V 76 mA
current ( open collector ) max. 50 mA !
Intrinsically safe only for EEx - version
HART (optional)
Drive coils
Temperature
Sensor PT100
Mass Flow Meter
Sensor RHM xx
Pick-up coil 1
Pick-up coil 2
1
2
3
4
5
6
7
8
9
TD+ > RD+ Pin 4
RD+ < TD+ Pin 2
TD - > RD - Pin 7
RD - < TD - Pin 9
SAFE AREA
RS 422/485 male 9 Pin
Aux. Input
Host
Source (5 VDC)
Sens (max. 5 VDC)
Ground
HAZARDOUS AREA
( only with option EEx ! )