Operation and Maintenance Manual – Steam sterilizers mod. PRATIKA
Rev. 01 del 03-01-2007 - C:\Documenti\Manuali di istruzione\PRATIKA GB-R01.doc
Pag. 16 di 27
5- It is advisable to purchase a printer with date stamp code M 9003 to file and record the sterilizing
cycles and any faults that may have occurred.
6 - There must always be trays and tray-holders in the chamber. Without these accessories, abnormal
pressures and temperatures could develop inside and these could cause the cycle to abort.
7 - It is advisable to place a phial containing culture spores in the chamber along with each load sterilized
in order to make sure that the sterilizing process is valid.
8 - When packeted materials are sterilized, it is very important to use support art. M9050. This support
can hold up to 20 packets (max. 3 instruments per packet). To use this accessory, it is essential to remove
the support and all trays from the chamber. The non-perforated tray may only be used to sterilize turbines
(handpieces) to prevent lubricant from being spread around the chamber. Only use perforated trays to
sterilize packeted instruments and always remove non-perforated ones from the chamber. If the packets
are not perfectly dry at the end of the sterilizing process, proceed with the supplementary drying cycle
with cycle (1-2) “special”.
7.3 Monthly inspection proposal
Fac simile of an inspection form
Sterilizer data
Brand
Model
Year of purchase
Observations
Periodical internal inspection
Date
Cycle length on
display
Cycle temperature
Cycle temperature
07/11/2000
20’
134°
VAPOR LINE
The utilized indicator is enclosed for each inspection.
Always use biological indicators. 3 types of integrator are indicated as an alternative.