SERENA Tecnico GB R-06 del 10.05.2019
00/ Per Bertani/ Manuale Tecnico Serena/ Serena
9
1.3.8 CONDUCTIVITY METER
T2 CHAMBER
T3 VAPORIZER
T4 BODY
T5 RESISTANCE
T6 RADIATOR
CONDUCTIVITY METER
Warning! By a certified instrument verify the quality of water in the tank.
Take a small amount of water from the blue grasp and proceed with the
verification.
THE FIRST THREE NUMBERS are used for calibration. Act with up / down arrow to change so that the
other three numbers correspond to the value found with the certificated instrument.
Values are expressed in µS.
CONDUCTIVITY METER
000 000 micro
1.4 INCREASE BAR
LOGO
SERIAL NUMBER
SW VERSION
CALIBRATION
INCREASE BAR
REFILLING
It detects the pressure (P) during vacuum phases.
INCREASE BAR
1050 mbar