SERENA Tecnico GB R-06 del 10.05.2019
00/ Per Bertani/ Manuale Tecnico Serena/ Serena
16
3
FUSES SCHEME
T1
: Transformer
F1
(250V T3,15A 5X20): It protects the power P. load, P. Vacuum, Fan, E.V. system, E.V. bacteriological filter, E.V. by
pass, E.V. vaporizer, E.V. drainage, E.V. load vaporizer pump, E.V. vacuum pump
F2
(250V T3,15A 5X20): It protects the power P. load, P. Vacuum, Fan, E.V. system, E.V. bacteriological filter, E.V. by
pass, E.V. vaporizer, E.V. drainage, E.V. load vaporizer pump, E.V. vacuum pump
F3
(250V T500mA 5X20): It protects the power of the display and the SD card low-voltage components
BT
: Battery (VARTA CR2032 3Vdc)
4
TECHNICAL EQUIPMENT
MINIMUM EQUIPMENT THAT THE TECHNICIAN MUST HAVE AT DISPOSAL FOR THE CALIBRATION:
CERTIFIED THERMOMETER WITH PROBE NO MORE THAN 3.2 mm
INSTRUMENT FOR MEASURING THE ENVIRONMENTAL PRESSURE
INSTRUMENT FOR MEASURING THE QUALITY 'OF WATER
TESTER TO VERIFY THE CONSUMPTION
TESTER TO CHECK THE SHORT CIRCUITS AND RESISTANCES
ALWAYS KEEP AN ORIGINAL PACKING AVAILABLE FOR EVENTUAL SHIPMENT OF THE STERILIZER
Note: All instruments must have the calibration certificate not expired!
INSTRUMENT FOR MEASURING THE
QUALITY 'OF WATER