![Reverberi M9010/100 Use And Maintenance Manual For The Technician Download Page 15](http://html1.mh-extra.com/html/reverberi/m9010-100/m9010-100_use-and-maintenance-manual-for-the-technician_1448321015.webp)
SERENA Tecnico GB R-06 del 10.05.2019
00/ Per Bertani/ Manuale Tecnico Serena/ Serena
15
2.
CHECK ACTUATOR:
A
=>
clean water above the minimum level
B
=>
clean water above the maximum level
C
=>
dirty water tank full
E
=> open door
D
=> door approached
F
=> closed door
EV VAC
=> 3-way electro valve
EV LOAD
=> vaporizer pump
EV DSCH
=> water drain electro valve
EV VAP
=> not connected
EV BYPS
=> bypass electro valve
EV FLT
=> bacteriological filter electro valve
EV RETE
(NETWORK) => purifier electro valve
VENT
=> condenser fan
P. VAC
=> vacuum pump
P. LOAD
=> charging pump clean water tank
R. CHAM
=> resistance chamber
R.VAP
=> vaporizer resistance
CLOSE DOOR
=> the engine is in the closed position and the micro are activated
OPEN DOOR
=> it opens and closes the door without considering the pressure. Warning! Before opening
the door verify that the pressure of the chamber is under vacuum or ambient pressure.
IMPORTANT NOTE: do not hold START when you do NOT hear the engine noise (ENGINE
UNPROTECTED).
If you are unable to open the door in any way, proceed as follows:
SETTINGS-> CONFIGURATION-> PASSWORD -> CHECK ACTUATOR -> OPEN DOOR -> START
press the button and release immediately when open the door!
In the process disappears F and appear the
E. Release the button just disappears the F and opens the door.
Do not hold pressed the button when
appears the E! Otherwise, the motor of the door BURNS (menu without protection!).
Use this procedure to open the door in case of EMERGENCY!
MENU WITHOUT PROTECTION!
Danger of burning the motor of the door!
Press ESC to exit and return to the main menu
Warning! If the controls do not respond properly, it is necessary to check the two fuses alongside the
electronic card. The correct positioning of the fuses is marked by a small noise of joints.
Small vibrations of the writings on the display indicate that there are anomalies in some component.
Water consumption: 1 lt every 2200 cycles -> 0,45 cc for 1 cycle