MAINTENANCE
CHIMIC DISINFECTION
COMPOSITION
:
The bath is equipped with an automatic system to determine the quantity of disinfectant.
The disinfection system includes :
- 1x 5 liters tank receiving the disinfectant product.
- 1X plunger with calibrated tube, located in the tank, and connected by a flexible going through the top of the
venturi (white plastic part).
- 1 disinfection shower.
ADDITION OF DISINFECTANT :
Dilution :
The mixing of water and disinfectant inside the disinfection shower is made in 2 stages :
First :
Manual dilution inside the tank.
Second :
The suction device (venturi) ensures a dosage of 10% of the aspirated product from the tank.
Exemple :
The disinfectant BFR 4.0 must be used with a final dilution of 0,4%.
1)
Manual dilution inside the tank at 4%, that is 200ml of BFR 4.0 for 4800ml of water.
2)
Complementary dilution of 10% ensured with
cold water
by the suction device (venturi).
Once the diluted product is added to the tank, any unused solution will need to be discarded and refreshed after one week.
CLEANING THE CONTROL PANEL
To clean the “rada sense” control panel, take the magnetic key located behind
the security door in front of the bath and pass it in front of the infra red of the control
panel to desactivate it. This will disable the sensors for 30 minutes or until the magnetic
key is reapplied. To reactivate the control panel, do the inverse operation.
CLEANING THE TUB & DOOR SEAL
Only use liquid products (liquid soap).
Don't use abrasive products like "scouring cream" to avoid scratches on the tub surface "Gelcoat".
Furthermore, to eliminate calcareous marks settled by water moving, we recommend to use the scaler (Gel Servex ).
Do not swill down the outside of the tub.
Do not use liquid soap and the scaler (Gel Servex) on the control panel elements;
Rada sense mixer and whirlpool control.
Open the door, clean and dry the door seal removing dirt. Wipe with a soft cloth
DISINFECTING THE TUB
•
Spray a suitable cleaner or disinfectant onto the interior tub surface.
•
Let the disinfectant activate for 2 minutes on the bath surface then rince off using the patient.
•
Thoroughly rinse the interior tub surfaces.
•
Dry the interior tub surfaces with a soft cloth.