6 Lubrication, Service and returning for maintenance
Pos : 10.2 /0025 War nhi nweis e/ W0003 WARNUNG R ei nigung Str omst oß neu @ 1\ mod_1236239978437_9.doc x @ 7686 @ @ 1
WARNING
Risk of a fatal electric shock
-
An electric shock can cause injuries in the form of burns and cardiac
arrhythmia, respiratory arrest or cardiac arrest.
•
Do not clean the blender under running water. Use only a cloth dampened
with water.
• Disconnect the power supply plug before cleaning the blender.
Pos : 10.3 /0025 War nhi nweis e/ W0012 WARNUNG EMBEDD ED R einigung Netzt trennen (PB100) @ 3 \ mod_1282722752283_9. doc x @ 22576 @ @ 1
WARNING
W0012
The device must always be switched off and disconnected from the mains before
any interventions for cleaning or servicing purposes.
Pos : 10.4 /0020 Übersc hrift en/1.1 Ü bersc hriften/ 1. 1 Ü berschrif ten BDA/11 R einigung @ 0 \ mod_1226909670239_9. doc x @ 3607 @ 2 @ 1
6.1 Lubrication
Fig. 16: Rear of the AT 100 XL
Fig 17: Lubrication points
The drive of the Abrasion Tester AT 100 XL has 2 lubrication
points which have to be lubricated regularly every 150 work
hours. The drive is located behind the cover (Fig. 16).
Only use lithium-based grease (
graphite free
) e.g. Shell
Gadus S2 V220 2 or BP Energrease LS-EP 2.