iBreeze BPAP Systems User Manual
2
Note: In either case above, it can only be determined by a competent physician whether
to use BPAP device.
User qualifications
The person operating the device by the instruction in the user
’s manual is referred as
the “user”. In contrast, a “patient” is the person receiving the therapy. Always perform all
the operating steps in accordance with these user’s manual.
Some lay person can receive specified professional training about how to use the
device including all related to accessories from your dealer or the manufacturer.
4 Safety Information
WARNINNG: Indicate the possibility of injury to the user or operator
.
CAUTION: Indicate the possibility of damage to the device.
WARNING:
This manual serves as a reference. The instructions in this manual are not intended
to supersede the health care professional’s instructions regarding the use of the
device.
This device is not intended for life-supporting device.
A mask should not be used unless the device is turned on, Otherwise, there is
danger of suffocation.
The device must be used only with the masks and accessories recommended by
Resvent. The masks and accessories are validated for use with Resvent devices.
The exhalation port(s) associated with the mask should never be blocked. The
device is intended to be used with special masks or connectors that have exhalation
ports to allow continuous flow of air out of the mask. When the device is turned on
and functioning properly, new air from the device flushes the exhaled air out through
the mask exhalation port. However, when the device is not operating, enough fresh
air will not be provided through the mask.
If you are using a full face mask (a mask covering both your mouth and your nose),
the mask must be equipped with a safety (entrainment) valve.
Sources of oxygen must be located more than 1 m from the equipment to avoid the
risk of fire and burns.
Do not use the device near a source of toxic or harmful vapors.
Do not use this device if the room temperature exceeds 35°C (95°F). If the device is
used at room temperatures warmer, than 35°C (95°F), the temperature of the airflow
may exceed 43°C (109°F). This could cause irritation or injury to your airway.
The device is not used exceed the specified temperature range.
Do not operate the device in direct sunlight or near a heating appliance because
these conditions can increase the temperature of the air coming out of the device.
Contact your health care professional if symptoms of sleep apnea recur.
If you notice any unexplained changes in the performance of this device, if it is
making unusual or harsh sounds, if it has been dropped or mishandled, if water is
spilled into the enclosure, or if the enclosure is broken, disconnect the power cord
and discontinue use. Contact your home care provider.
Repairs and adjustments must be performed by Resvent authorized service
personnel only. Unauthorized service could cause injury, invalidate the warranty, or
result in costly damage.
Do not use any accessories, detachable parts, and materials not recommended by
Resvent. Incompatible parts or accessories can result in degraded performance.