iBreeze BPAP Systems User Manual
30
16 EMC Declaration
Essential Performance
:
BPAP mode: IPAP 12 cmH2O, EPAP 10 cmH2O.
The BPAP system is intended for use in the specified electromagnetic environment
listed in form B-1, B-2 and B-3. The customer or the user of BPAP system should
assure that it is used in such an environment as described below.
The BPAP system is intended for use in an electromagnetic environment in which
radiated RF disturbance are controlled. The customer or the user of BPAP system can
help prevent electromagnetic interference by maintaining a minimum distance between
portable/mobile RF communications equipment (transmitters) and the BPAP system (as
recommended in form B-3) , according to the maximum output power of the
communications equipment.
Form B-1Guidance and Declaration - Electromagnetic Emissions
Emissions test
Compliance
Electromagnetic environment -
guidance
Radiated emissions
CISPR11
Group 1
The BPAP system uses RF
energy only for its internal
function. Therefore, its RF
emissions are very low and are
not likely to cause any
interference in nearby electronic
equipment.
Conducted emissions
CISPR11
Class B
The BPAP system is suitable for
use in all establishments,
including domestic
establishments and those
directly connected to the public
low-voltage power supply
network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC61000-3-3
Complies