Summary of Contents for iBreeze 25A

Page 1: ......

Page 2: ...formation 2 5 Symbol 6 6 System Contents 7 7 System Overview 7 8 Navigating the Device Screens 11 9 Operation 12 10 Device Alert 21 11 Troubleshooting 24 12 Cleaning and Maintaining 25 13 Maintenance...

Page 3: ...A and Chronic Obstructive Pulmonary Diseases COPD in spontaneously breathing patients weighting over 30kg 66lbs it is for use in the home hospital or institutional environment The system is also inten...

Page 4: ...ly new air from the device flushes the exhaled air out through the mask exhalation port However when the device is not operating enough fresh air will not be provided through the mask If you are using...

Page 5: ...ter frequently for increased resistance and blockage to ensure the delivery of the therapeutic pressure Please check whether there is water in the device before use The maximum fill level is 290 ml Fa...

Page 6: ...und the device for the system to work properly Make sure the gas intake port on the side of the device is not blocked by bedding curtains or other items Air must flow freely around the device for the...

Page 7: ...e device in home environment In case you feel comfortless when you contact the device You shall stop to use the device and contact your supplier immediately Because it may cause allergic reactions Ple...

Page 8: ...th the provisions of the Council Directive 93 42 EEC concerning medical devices and fulfils the essential requirements of Annex I of this directive European Authorized Representative Dispose according...

Page 9: ...clable materials 6 System Contents Your BPAP system may include the following items BPAP Device Power Adapter Power Cord SD Card Flexible Tube 19 mm or 15 mm tube Travel Bag User Manual NOTE if any of...

Page 10: ...lower than your sleeping position Make sure the device is away from any heating or cooling equipment e g forced air vents radiators air conditioners Note When positioning the device make sure that th...

Page 11: ...to discontinue use and replace if damaged Please first check the DC connector could be used normal and not breaking free Never place the power cord around the neck Do not use any small parts to fix p...

Page 12: ...use low resistance and bacteria filtered cotton between the equipment and the tube Note The selected mask breathing tube shall meet the performance requirements with good stability The user should che...

Page 13: ...Interface UI on this device allows you to adjust the device settings and view information about your therapy The UI is comprised of the display screen and the control dial Rotate the control dial in e...

Page 14: ...e sure that no air is leaking from your mask If necessary adjust the mask and headgear until the air leak stops 5 If you are using the device in a bed with a headboard try placing the tubing over the...

Page 15: ...n 6 Mask Fit Mask fit feature allows you to check the fit of your mask prior to starting therapy This is done by measuring the amount of leak 7 Comfort Press to enter Comfort setting interface 8 Setti...

Page 16: ...level is available when the humidity function is on 4 Background breathing dynamic diagram 5 The current therapy pressure 6 The therapy parameters monitoring during therapy 7 Displays the oxygen satur...

Page 17: ...Real time monitoring parameters 3 Switch to the patient therapy clean interface 4 Flow real time waveform 5 Pressure real time waveform 6 Work status icon bar 7 Display the alarm messages 9 2 3 Patie...

Page 18: ...cmH2O Auto Ramp Turn Auto Ramp function on or off When Auto Ramp is on Sleep state can be identified automatically sleep begins to start Ramp function awake do not start ramp function Option On OFF D...

Page 19: ...ace language Default English Chinese Default English Energy Saving Set the Energy Saving function on or off When Energy Saving is on the screen will be turned off automatically if there is no operatio...

Page 20: ...RS Set the GPRS function on or off Option ON OFF Default OFF Reminder Set the Reminder type and time You can turn on or off the different type of reminders including Mask Filter Water tank Tube Clinic...

Page 21: ...nths 6 months 1 year Usage Number of hours the device has been used in the selected period AHI The average apnea hypopnea index in the selected period Mask fit Use three stars for rating the leakage o...

Page 22: ...re provider may ask you to send the SD card to them for evaluation To remove the SD card 1 Stop therapy and open the SD card cover 2 Push in the SD card to release it Remove the SD card from the devic...

Page 23: ...ge is too high or too low Use the Power Adapter provided by Resvent SD card write read underway do not remove the SD card do not cut off the power Notification 1 Insert SD card during data synchroniza...

Page 24: ...e the usage time exceed the water tank setting service life Click Confirm replace the water tank The filter expired please replace Notification In standby state the usage time exceed the filter settin...

Page 25: ...iration tube or adjust therapy parameter setting Low Expiratory Pressure Alert 0 During breath cycle monitored pressure is lower than the set pressure threshold Check the respiration tube or adjust th...

Page 26: ...ing instructions or use the Mask Fit function to check your mask fit and seal I am getting a dry or blocked nose Humidity level may be set too low Adjust the Humidity Level I am getting droplets of wa...

Page 27: ...aning the Device WARNING To avoid electrical shock always unplug the power cord from the wall outlet before cleaning the device Do not immerse the device in any fluids 1 Unplug the device Wipe the out...

Page 28: ...ling Please follow the below steps for packing 1 Remove the water tank from the device and pour out all water 2 Install the water tank back on the device 3 Put the device and accessories in the travel...

Page 29: ...1 1 Storage Information Store the device under the prescribed ambient conditions 14 1 2 Storing the therapy device Switch off the therapy device Disconnect the therapy device from the power supply Cl...

Page 30: ...l equipment part 1 2 general requirements for basic safety and essential performance collateral standard electromagnetic disturbances requirements and tests ISO 80601 2 70 2015 Medical electrical equi...

Page 31: ...e 4 30 cmH2O Max Single Fault Steady Pressure 40 cmH2O Pressure Control Accuracy 0 5 cmH2O Dynamic Pressure Variation 1 cmH2O Pressure monitoring Accuracy 2 full scale reading 4 actual reading Flow Fl...

Page 32: ...ent transmitters and the BPAP system as recommended in form B 3 according to the maximum output power of the communications equipment Form B 1Guidance and Declaration Electromagnetic Emissions Emissio...

Page 33: ...ty should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC61000 4 11 0 70 0 of UT 0 for 0 5 cycle Mains p...

Page 34: ...e P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance Field strengths from fixed RF transmitters a...

Page 35: ...esvent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this manual Contents of this manual are subjec...

Page 36: ......

Page 37: ...Road Xinan Street Baoan District Shenzhen 518100 P R China Factory Address Room 802 Building 2 Gaoxinqi Industrial Park phase II Liuxian No 1 Road Xinanstreet Baoan District Shenzhen 518100 P R China...