To adjust the Pause Time, select PAUSE TIME from
the
Setup Menu 1
screen described on page
21
.
Use the DOWN and UP buttons to set the Pause Time
from OFF to 10 minutes in half minute increments.
When the desired setting is shown, select SAVE.
Select BACK to return to
Setup Menu 1
without saving changes.
This sets the maximum pressure allowed for all programs and programming. To adjust
the Pressure Limit, select PRESS LIMIT from the
Setup Menu 2
screen described on
page
21
. Use the DOWN and UP buttons to set the system pressure limit. When the
desired setting is shown, select SAVE.
Select BACK to return to
Setup Menu 2
without saving changes.
This sets the pressure increment used when adjusting pressure for all running sessions
and session programming.
To adjust the Pressure Increment, select PRESS INC from the
Setup Menu 2
screen
described on page
21.
Use the DOWN and UP buttons to set the pressure increment
to 5% or 10%. When the desired setting is shown, select SAVE.
Select BACK to return to
Setup Menu 2
without saving changes
Pressure Increment:
Reset Defaults and Clear Average Use:
This allows the user to reset settings to original factory parameters. There are three
reset options. All defaults (ALL DFLTS) resets all programs and setup menu variables.
Setup defaults (SETUP DFLTS) resets only setup menu variables. CLEAR USE resets the
session count and average use information to zero.
To reset settings to original factory defaults, select NEXT then RESET DFLTS from the
Setup Menu 2
screen described on page
21.
The RESET DFLTS menu will display the
three reset options to select: ALL DFLTS, SETUP DFLTS, and CLEAR USE. Selecting a
reset option will display a confirmation screen to continue. Select YES to confirm or
select NO to cancel.
NEXT
PRES LIMIT
(i)
SETUP MENU
PRESS:
(i)
SETUP MENU
PRESS:
Pressure Limit:
NEXT
PRES INC
NEXT
RESET DFLTS
(i)
SETUP MENU
PRESS:
P A U S E T I M E : 1 . 5 m
B A C K D O W N U P S A V E
SYSTEM UTILITIES (cont’d)