Chapter 9
9-42
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Humidifier
Receptacle Rotation
Rotate the humidifier AC receptacle 180 degrees if necessary to allow right-
angle humidifier power cords to connect to the receptacle. Follow these steps
to rotate the receptacle (Figure 9-48, Figure 9-49, Figure 9-50, Figure 9-51):
1. Remove the top enclosure (see “Top Enclosure” on page 9-4) to allow
access to the AC distribution panel.
2. From inside the ventilator, unscrew the blue neutral wire from the
humidifier circuit breaker.
Figure 9-48: Removing the Blue Neutral Wire
Blue neutral wire
Remove this
end first
Humidifier
circuit
breaker