Chapter 9
9-44
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Figure 9-51: Rotated Humidifier AC Receptacle
8. Connect the free end of the replacement neutral wire to the humidifier
circuit breaker and screw into place, with the replacement wire in the
same orientation as the other circuit breaker connections.
9. Once installed, plug in a humidifier or measure AC voltage to verify
that the AC receptacle operates correctly. Perform the electrical safety
test and EST as described in Chapter 8.
Printed Circuit Boards
(Daughter PCBs)
(Except Main PCB)
Follow these steps to remove the daughter PCBs (Figure 9-52, Figure 9-56).
Reverse removal steps to install.
NOTE:
NOTE: The main PCB must be installed before reinstalling the other
PCBs.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Loosen the three captive screws and remove the screw holding the
PCB shroud, then remove the cover.
3. Pull up on the top corners of the daughter PCB you want to remove,
using a slight rocking motion as you pull.
Before
rotation
After
rotation
Rotated humidifier outlet allows connection of right angle power cords