Chapter 7
Testing and Calibration
1062788
BiPAP Vision Service Manual © Respironics, Inc.
7-7
7.2 Service Laptop
Setup
This section describes how to set up a service laptop for testing and
troubleshooting the BiPAP Vision. The ventilator includes all necessary
software to work with a service laptop, and does not require additional
software.
7.2.1
Selecting the Test Cable
Select the cable(s) for service laptop setup (Table 7-3) according to the serial
number and upgrade status of the ventilator:
Table 7-3: Test Cables for Service Laptop
Ventilator serial number/upgrade
Required cables
Ventilators s/n 105999 and below, no
upgrades (original MCS, PAS, and
DCS boards installed)
•
Test cable
582161
Ventilators s/n 105999 and below,
upgraded (new PC and MC boards
installed)
•
Test cable
1004823
•
Ribbon cable
1004699
•
Standard RS-232 serial cable
Ventilators s/n 105999 and below
(upgraded) with original DCS board
installed
•
Test cable
1004823
•
Test cable
582161
connected to
DCS board
•
Ribbon cable
1004699
•
Standard RS-232 serial cable
Ventilators s/n 106000 and above
•
Test cable
1004823
•
Standard RS-232 serial cable