42
XII.
Troubleshooting Guide
Symptom
Possible Cause
Possible Solution
920 will not power
up
Batteries are
completely depleted
Replace all six batteries
of the 920
Incorrect battery
installation
Verify battery sequence.
Refer to Figure I:
Replacing Batteries in
the 920
Battery door is missing
metal contact
Replace the battery
door
Numeric displays
are blinking at
once per second
Battery voltage is low
Replace all six batteries
of the 920
Incorrect battery
installation
Verify battery sequence.
Refer to Figure I:
Replacing Batteries in
the 920
Minus sign
appears in SpO
2
display
Sensor fault exists.
Sensor may have
become dislodged
from 920 or from the
patient
Verify sensor is
connected to the 920
and the patient correctly;
Try a new sensor if
condition persists
Segments of
SpO
2
or
♥
display
are missing
Defective LED displays Displayed values may
not be reliable;
discontinue use of 920
Displayed pulse
rate does not
correlate to pulse
rate displayed on
ECG monitor
Excessive motion at
sensor site may be
prohibiting the 920
from acquiring a
consistent pulse signal
Eliminate or reduce
cause of motion artifact
or reposition sensor to
new sensor site where
motion is not present
Patient may have an
arrhythmia resulting in
some heart beats that
do not yield a perfusion
signal at sensor sight
Examine the patient:
Condition may persist
even though both
monitors are functioning
properly if patient's
arrhythmia persists
Non- Respironics
sensor is being used
Replace sensor with a
Respironics sensor
ECG monitor may not
be functioning properly
Examine the patient:
Replace ECG monitor
or refer to operator's
manual for ECG monitor
Summary of Contents for 920
Page 1: ...English Instruction and Service Manual Models 920 and 920M Hand Held Pulse Oximeters ...
Page 2: ......
Page 4: ...iv ...
Page 8: ...viii ...