Finger Clip Sensors
NOTE:
For the best results, secure the sensor cable
independently from the sensor, preferably
around the base of the finger. Make sure that
the tape securing the cable does not restrict the
blood flow.
NOTE:
If the patient's finger is too small or the sensor
is not adjusted properly, light may bypass the
tissue and result in SpO
2
inaccuracies.
1. Adult Articulated Finger Clip Sensor
Insert finger into the sensor until the end of the finger reaches
the finger stop. Ensure that long fingernails are not interfering
with proper finger position. The thumb is specifically not
recommended for use with the Adult Articulated Finger Clip
Sensor.
Figure VIII: Adult Articulated Finger Clip Sensor.
2. Pediatric Finger Clip Sensor
Insert finger into the sensor until the end of the finger reaches
the finger stop. Ensure that long fingernails are not interfering
with proper finger position. The thumb is specifically not
recommended for use with the Pediatric Finger Clip Sensor.
Figure IX: Pediatric Finger Clip Sensor.