Technical Specifications
The manufacturer reserves the right to change these specifications without notice.
Symbols which appear on the device
Attention, consult accompanying documents;
Drip proof;
Type CF equipment;
Dangerous voltage;
Class II equipment;
Start/Stop.
Operating pressure range
4 to 20 cm H
2
O
Dimensions (H x W x D)
4.6" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
Weight
3.1 lb (1.4 kg)
Operating Temperature
+41°F to +95°F (+5°C to +36°C)
Power Supply
•
Input range for S8 Series II with HumidAire 3i: 100–240V, 50/60Hz;
110V, 400Hz; 2.5A <140VA (110W) (maximum power consumption).
Instantaneous peak power consumption <340VA.
•
Refer to the DC-12 converter instructions for DC ratings. Only use
the ResMed DC-12 converter for DC input.
•
Typical power consumption at 20 cm H
2
O is 39VA or 20.7W.
Operating Humidity
10%–95% non-condensing
Storage and Transport
Temperature
-5°F to -140°F (-20°C to +60°C)
Storage and Transport
Humidity
10%–95% non-condensing
Operating Altitude
Sea level to 8500’ (2591 m)
Electromagnetic
Compatibility
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial and light industry environments. The electromagnetic
compatibility tables for these ResMed devices can be found on
www.resmed.com
, on the
Products
page under
Service and
Support
. Click on the PDF file for your device.
Air Filter
Two-layered, powder-bonded, polyester non-woven fibre
Air Tubing
Flexible plastic, 6’ 6" (2 m)
Air Outlet
The 22 mm conical air outlet complies with EN 1281-1
IEC 60601-1 Classifications
Class II (double insulation), Type CF
Supplemental Oxygen
Recommended maximum supplemental oxygen flow: 4 L/min
Housing Construction
Flame retardant engineering thermoplastic
Noise Level
Radiated sound pressure is measured at <24±2 dBA at 10 cm H
2
O and
3’3.36” (1 m)
according to ISO17510-1