Use filters and reading scale
Analyse data
Signals
High-pass filter
Low-pass filter
Usual reading scale
EEG (C3, C4, F3, F4, O1, O2, M1 and M2)
0.3 Hz
35 Hz
70 μv/cm
EOG (E1 and E2)
0.3 Hz
35 Hz
100 μv/cm
EMG (Chin 1-F and 1-2, left/right leg)
10 Hz
100 Hz
100 μv/cm
EKG
0.3 Hz
70 Hz
Automatic
Right click on the ExG signal and choose the corresponding filter in the menu
Filter > Filter presets.
You can manually adjust filter values in the menu
Filter out frequencies below/above
.