17
Patient hook-up
PLM snap-on lead wire
Grey (PN 7073438)
Electrode site:
Connection to the Nox A1:
PLM snap-on lead wire
Black (PN 7073439)
4 disposable adhesive electrodes
(PN 631526)
Scrub the patient’s skin with NuPrep before placing
the adhesive electrodes.
Electrodes must be placed on the tibialis anterior muscle.
If needed, secure the electrodes with tape and/or a Surgilast net.
•
Place the PLM cables
Equipment needed:
Connect the black cable (left leg) to the LM1 bipolar input and
the grey cable (right leg) to the LM2 bipolar input.